Percent Oxygen Analyzer

Installation

 

 

 

The alarm relay circuits are designed for failsafe operation, meaning the relays are energized during normal operation. If power fails the relays de-energize (alarms activated).

The contact connections are indicated diagrammatically on the rear panel as Normally Closed, Common, and Normally Open. Figure 3-3 explains how these act in failsafe operation.

Alarm 1 and Alarm 2 are configured as HI and LO respectively. A HI alarm will activate when concentration is above threshold, while a LO alarm will activate when concentration is below threshold.

Figure 3-3: Contact ID for FAILSAFE Relay Operation

The specific descriptions for each type of alarm are as follows:

Alarm #1

Programmable as high alarm. Can be set anywhere

 

between 0 and 100 % but no lower than Alarm #2.

Alarm #2

Programmable as low alarm. Can be set anywhere

 

from 0 to 100 % but no higher than Alarm #1.

Sensor Fail

Actuates when the output of the sensor falls below a

 

factory programmed acceptable level or the

 

thermocouple is open.

CAUTION:

THERE COULD BE HAZARDOUS VOLTAGE AT THE

 

ALARMS TERMINALS, EVEN WHEN POWER IS

 

REMOVED FROM THE INSTRUMENT.

Teledyne Analytical Instruments

27

Page 27
Image 27
Teledyne 9060L, Percent Oxygen Analyzer operating instructions Alarm #2