Percent Oxygen Analyzer

Maintenance

 

 

 

Maintenance

Aside from normal cleaning and checking for leaks at the probe connection, the Model 9060L should not require any maintenance beyond replacement of a blown fuse. Routine maintenance includes occasional recalibration, as described in Chapter 4, Operation.

5.1 Replacing the Fuse

CAUTION: REMOVE POWER TO UNIT BEFORE REPLACING THE FUSE.

When a fuse blows, check first to determine the cause, then replace the fuse using the following procedure:

1.Disconnect the AC power and place the power switch located on the rear panel in the O position. Remove the power cord from the receptacle.

2.The fuse receptacle is located in the power cord receptacle assembly in the upper left-hand corner of the rear panel. See Figure 5-1.

Figure 5-1: AC Fuse Replacement

Teledyne Analytical Instruments

41

Page 41
Image 41
Teledyne 9060L operating instructions Replacing the Fuse, Percent Oxygen Analyzer Maintenance