Percent Oxygen Analyzer

Appendix

 

 

 

Appendix

A.1 Specifications

Ranges: 0-3 % and 0-10 % oxygen (Standard Ranges), and 0-25 % (nominal) Calibration Range. User selectable % Ranges can be set between 1% and 100 % oxygen .

Signal Output: Voltage: 0–10 VDC, negative ground

Current: 4-20 mA, negative ground

Range ID: 0-10 VDC.

Display: Light emitting diode display.

Alarms: Two customer selectable fully adjustable

HI and LO alarms.

One sensor failure alarm.

Alarm relays form "C" contacts, dry contacts rated at 2A for 30VDC, 1A for 115VAC.

System Operating Temp: 0-50 °C

Accuracy: ±2% of full scale at constant temperature ±5% of full scale through operating temp. range once temp. equilibrium is reached. (At 3 % and higher user defined ranges.)

Response Time: 90% in less than 4 seconds at 25 °C.

System Power Requirements: 120 VAC, 50/60Hz (optional 220VAC), 110W.

System Enclosure: Panel mount. Dimensions (Approx.) 7.0 W x 6.75 H x 4.26 D.

Sensor Type: Zirconia, solid state

Teledyne Analytical Instruments

43

Page 43
Image 43
Teledyne 9060L, Percent Oxygen Analyzer operating instructions Appendix, Specifications