Appendix

Model 9060L

 

 

 

A.2 Probe Specifications

Application: Combustion flue below 700°C (1290°F), above 700°C (1290°F) 900°C (1650°F) — no contaminants.

With contaminants see notes 1 &2

Temperature Range:

0-900°C

800-1400°C

800-1200°C

 

(32-1650°F)

(1470-2550°F)

(1470-2190°F)

Length:

250-1000 mm 500-1160 mm 457-1219 mm

 

(10” to 40”)

(20” to 46”)

(18” to 48”)

Process Connection: 1 1/2" BSP 3/4" BSP 1" BSP or equivalent

NPT

Electrical Connection: Weatherproof plug-in connector or optional screw terminals. Plug connector supplied with the cable. Ex d heads have screw terminals.

Cable: Order a specific length with the analyzer except for hazardous installations where the cable is supplied by the customer.

Heater: Yes

Flue Gas Thermocouple: K

Response Time: Typically < 4 secs.

Head Temperature: 125°C (250°F)

Reference Gas: Ambient air 50 to 500 cc/min (6 to 60 scfm). Pump supplied with analyzer

Ref Air Connection: 1/4" NPT integral air line in probe cable. Barbed fitting to 3/16" ID PVC tube.

Filter: Removable sintered stainless steel particulate filter, 15 micron.

Calibration Gas Connection: 1/8" NPT female 1/8" NPT female 1/8"

NPT female

Weight : 0.6 kg (1.32 lbs.)

Teledyne Analytical Instruments

44

Page 44
Image 44
Teledyne Percent Oxygen Analyzer operating instructions Probe Specifications, Appendix Model 9060L