Teledyne API – Model T300/T300M CO Analyzer Introduction, Features and Options
29
Option Option
Number Description/Notes Reference
Special Features Built in features, software activated
N/A
Maintenance Mode Switch, located inside the instrument, places the
analyzer in maintenance mode where it can continue sampling, yet
ignore calibration, diagnostic, and reset instrument commands. This
feature is of particular use for instruments connected to Multidrop or
Hessen protocol networks.
Call Customer Service for activation.
N/A
N/A
Second Language Switch activates an alternate set of display
messages in a language other than the instrument’s default language.
Call Customer Service for a specially programmed Disk on Module containing
the second language.
N/A
N/A
Dilution Ratio Option allows the user to compensate for diluted
sample gas, such as in continuous emission monitoring (CEM) where
the quality of gas in a smoke stack is being tested and the sampling
method used to remove the gas from the stack dilutes the gas.
Call Customer Service for activation.
Sections 3.4.4.2
and 5.4.5
06864B DCN6314