Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840) APPENDIX A-2: Setup Variables For Serial I/O
A-11
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
ET_TARGET_DET mV 4375 0–5000
Target detector reading during
electrical test.
ET_TARGET_CONC PPM 40,
400 3, 8
1–9999.99 Target concentration durin g
electrical test.
ET_CONC_RANGE Conc. 50,
5000 3, 8
0.1–50000 D/A concentration range duri ng
electrical test.
STD_TEMP ºK 321 1–500
Standard temperature for
temperature compensation.
STD_PRESS "Hg 28.5,
28.7 8,
28.8 12, 18,
28.1 4
1–50 Standard press ure for pressure
compensation.
48 BENCH_SET ºC
Warnings:
43–53
0–100 Optical bench temperature set
point and warning limits.
68,
62 19,23
WHEEL_SET ºC
Warnings:
63–73,
57–67 19,23
0–100 Wheel temperature set point and
warning limits.
50 O2_CELL_SET 14 ºC
Warnings:
45–55
30–70 O2 sensor cell temperature set
point and warning limits.
STD_O2_CELL_TEMP 14 ºK 323 1–500 Standard O2 cell temperature for
temperature compensation.
ZERO_APPLY_IN_CAL 5 ON OFF, ON
ON applies auto-reference offset
and dilution factor during
zero/span calibration;
OFF disables both.
(Only applicable if
ZERO_ENABLE is ON.)
ZERO_DWELL 3, 5, 8 Seconds,
Minutes 5
7,
3 5
1–60,
1–30 5
Dwell time after closing or
opening zero scrubber valve.
ZERO_SAMPLES 3, 5, 8 Samples 15,
750 5,
1000 19
1–1000 Number of zero samples to
average.
ZERO_FILT_SIZE 3, 5, 8 Samples 5,
1 5
1–100 Auto-zero offset moving average
filter size.
ZERO_LIMIT 3, 5, 8 Ratio 1.2,
1.15 3, 8,
1 5
0–5 Minimum auto-zero ratio allo wed;
must be greater than this value
to be valid.
ZERO_CAL 3, 5, 8 Ratio 1.18 0.5–5 Calibrated auto-zero ratio.