Teledyne API – Model T300/T300M CO Analyzer Specifications and Approvals
33
Table 2-3: CO2 Sensor Option Specifications
2.2. EPA EQUIVALENCY DESIGNATION

Note T300M: EPA equivalency does not apply to this model.

Teledyne API’s Model T300, Carbon Monoxide Analyzer, is designated as Reference

Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under

the following conditions:

Range: Any range from 10 ppm to 50 ppm
Ambient temperature range of 10 to 40C
Sample filter: Equipped with 5-micron PTFE filter element in the internal filter
assembly
Software settings:
Dilution factor 1.0
AutoCal ON or OFF
Dynamic Zero ON or OFF
Dynamic Span OFF
Dual Range ON or OFF
Auto Range ON or OFF
Temp/Pres Compensation ON
Parameter Description
Ranges 0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported.
Zero Noise1 <0.02% CO2
Lower Detectable Limit2 <0.04% CO2
Zero Drift (24 hours) <± 0.02% CO2
Zero Drift (7 days) <± 0.05% CO2
Span Noise1 <± 0.1% CO2
Span Drift (7 days) <± 0.1% CO2
Accuracy <± (0.02% CO2 + 2% of reading)
Linearity <± 0.1% CO2
Temperature Coefficient <± 0.01% CO2 /°C
Rise and Fall Time <60 seconds to 95%
1 As defined by the USEPA
2 Defined as twice the zero noise level by the USEPA
06864B DCN6314