Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840) APPENDIX A-2: Setup Variables For Serial I/O
A-17
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
BENCH_DERIV — 2
0 19,23
0–100 100V optical bench temperature
PID derivative coefficient.
BENCH_PROP2 1/ºC 5
0.75 19,23
0–100 200V optical bench temperature
PID proportional coefficient.
Proportional band is the
reciprocal of this setting.
BENCH_INTEG2 — 0.5
0.75 19,23
0–100 200V optical bench temperature
PID integral coefficient.
BENCH_DERIV2 — 2
0 19,23
0–100 200V optical bench temperature
PID derivative coefficient.
WHEEL_CYCLE Seconds 4
2 4,9,12,18
8 19,23
0.5–30 Wheel temperature control cycle
period.
WHEEL_PROP 1/ºC 1
0.3 19,23
0–100 100V wheel temperature PID
proportional coefficient.
Proportional band is the
reciprocal of this setting.
WHEEL_INTEG — 0.135
0.035 4,9,12,18
0.06 19,23
0–100 100V wheel temperature PID
integral coefficient.
WHEEL_DERIV — 2
0 19,23
0–100 100V wheel temperature PID
derivative coefficient.
WHEEL_PROP2 1/ºC 1
0.1 19,23
0–100 200V wheel temperature PID
proportional coefficient.
Proportional band is the
reciprocal of this setting.
WHEEL_INTEG2 — 0.135
0.035 4,9,12,18
0.01 19,23
0–100 200V wheel temperature PID
integral coefficient.
WHEEL_DERIV2 — 2
0 19,23
0–100 200V wheel temperature PID
derivative coefficient.
O2_CELL_CYCLE 14 Seconds 10 0.5–30
O2 cell temperature control cycle
period.
O2_CELL_PROP 14 — 1 0–10
O2 cell PID temperature control
proportional coefficient.
O2_CELL_INTEG 14 — 0.1 0–10
O2 cell PID temperature control
integral coefficient.
O2_CELL_DERIV 14 — 0 (disabled) 0–10
O2 cell PID temperature control
derivative coefficient.
BOX_TEMP_GAIN PPB/DegC 0,
5 9
0–100 Gain factor for box temperature
compensation of concentration.
TPC_ENABLE — ON OFF, ON
ON enables temperature/
pressure compensation; OFF
disables it.
CONC_LIN_ENABLE — ON OFF, ON
ON enables concentration
linearization; OFF disables it.
STAT_REP_PERIOD 17 Seconds 1 0.5–120
TAI protocol status message
report period.