Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840) APPENDIX A-2: Setup Variables For Serial I/O
A-15
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
TEST_CHAN_ID — NONE
0 NONE,
CO
MEASURE,
CO
REFERENC
E,
VACUUM
PRESSURE,
SAMPLE
PRESSURE,
SAMPLE
FLOW,
SAMPLE
TEMP,
BENCH
TEMP,
WHEEL
TEMP,
O2 CELL
TEMP 14,
CHASSIS
TEMP,
PHT DRIVE,
TEMP4 5
Diagnostic analog output ID.
REMOTE_CAL_MODE — LOW
0 LOW,
HIGH,
CO2 10,
O2 14
CO range or other gas to
calibrate during contact closure
or Hessen calibration.
PASS_ENABLE — OFF ON, OFF
ON enables passwords; OFF
disables them.
STABIL_FREQ Seconds 10
120 19, 23
1–300 Stability measurement sampling
frequency.
STABIL_SAMPLES Samples 25 2–40
Number of samples in
concentration stability reading.
2500 PHOTO_TEMP_SET mV
Warnings:
250–4750
0–5000 Photometer temperature warning
limits. Set point is not used.
29.92 SAMP_PRESS_SET In-Hg
Warnings:
15–32
0–100 Sample pressure warning limits.
Set point is not used.
06864B DCN6314