Overview of Operating Modes Teledyne API – Model T300/T300M CO Analyzer 92 Table 4-2: Test Functions Defined
PARAMETER DISPLAY TITLE UNITS MEANING
Stability STABIL PPB3, PPM
UGM3, MGM
Standard deviation of CO concentration readings. Data points are
recorded every ten seconds using the last 25 data points. This
function can be reset to show O2 or CO2 stability in instruments with
those sensor options installed.
Range
RANGE
RANGE11
RANGE21
PPB, PPM,
UGM, MGM
The full scale limit at which the reporting range of the analyzer is
currently set. THIS IS NOT the Physical Range of the instrument.
See Section 5.4.1 for more information.
O2 Range 1 O2 RANGE % The range setting for the optional O2 Sensor.
CO2 Range2 CO2 RANGE % The range setting for the optional CO2 Sensor.
CO Measure CO MEAS MV The demodulat ed, peak IR detector output during the measure
portion of the GFC Wheel cycle.
CO Reference CO REF MV The demodulated, peak IR detector output during the reference
portion of the GFC Wheel cycle.
Measurement /
Reference Ratio MR Ratio -
The result of CO MEAS divided by CO REF. This ratio is the
primary value used to compute CO concentration. The value
displayed is not linearized.
Sample Pressure PRES In-Hg-A The absolute pressure of the Sample gas as measured by a
pressure sensor located inside the sample chamber.
Sample Flow SAMPLE FL cm3/min Sample mass flow rate as measured by the flow rate sensor in the
sample gas stream.
Sample
Temperature SAMP TEMP C The temperature of the gas inside the sample chamber.
Bench
Temperature BENCH TEMP C Optical bench temperature.
Wheel
Temperature WHEEL TEMP C GFC Wheel temperature.
Box Temperature BOX TEMP C The temperature inside the analyzer chassis.
O2 Cell
Temperature1 O2 CELL TEMP3 C The current temperature of the O2 sensor measurement cell.
Photo-detector
Temp. Control
Voltage
PHT DRIVE mV The drive voltage being supplied to the thermoel ectric coolers of the
IR photo-detector by the sync/demod Board.
Slope SLOPE - The sensitivity of the instrument as calculated during the last
calibration activity.
Offset OFFSET - The overall offset of the instrument as calculated during the last
calibration activity.
O2 Sensor
Slope 1 O2 SLOPE - O2 slope, computed during zero/span calibration.
O2 Sensor Offset
1 O2 OFFSET - O2 offset, computed during zero/span calibration.
CO2 Sensor
Slope2 CO2 SLOPE - CO2 slope, computed during zero/span calibration.
CO2 Sensor
Offset 2 CO2 OFFSET - CO2 offset, computed during zero/span calibration.
Current Time TIME - The current time. This is used to create a time stamp on DAS
readings, and by the AUTOCAL feature to trigger calibration events.
1 Only appears when the optional O2 sensor is installed.
2 Only appears when the optional CO2 sensor is installed.
3 Only available on the T300.