Teledyne API – Model T300/T300M CO Analyzer Getting Started

51

Table 3-6: Status Output Signals

REAR PANEL
LABEL
STATUS
DEFINITION CONDITION
1 SYST EM OK ON if no faults are present.
2 CONC VALID
OFF any time the HOLD OFF feature is active, such as during calibration or when
other faults exist possibly invalidating the current concentration measurement
(example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.
3 HIGH RANGE
ON if unit is in high range of either the DUAL or AUTO range modes.
4 ZERO CAL
ON whenever the instrument’s ZERO point is being calibrated.
5 SPAN CAL
ON whenever the instrument’s SPAN point is being calibrated.
6 DIAG MODE
ON whenever the instrument is in DIAGNOSTIC mode.
7 CO2 CAL
If this analyzer is equipped with an optional CO2 sensor, this Output is ON when that
sensor is in calibration mode.
Otherwise this output is unused.
8 O2 CAL
If this analyzer is equipped with an optional O2 sensor, this Output is ON when that
sensor is in calibration mode.
Otherwise this output is unused.
D EMITTER BUS The emitters of the transistors on pins 1- 8 are bussed together.
SPARE
+ DC POWER + 5 VDC, 300 mA source maximum.

Digital Ground The ground level from the analyzer’s internal DC po wer supplies.
3.3.1.6. CONNECTING THE CONTROL INPUTS To remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel. There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is the most convenient method. However, if full isolation is required, an external 5 VDC power supply should be used.