LF410/LF600 LF410/LF602
Fluid conductivity: 5µS/cm minimum
Fluid temperature:
Note : 120°C (248°F) above is separate type
(14 to 248°F)
Ambient temperature:
Without cFMus Approval:
With cFMus Approval:
Structure: NEMA 4X (IP 67) Watertight
Power consumption:
17W(27VA) or less
19W(29VA) or less (with PROFIBUS)
Conformance to European Community Directives:
EMC directive 89/336/EEC
The low voltage 93/68/EEC
PED 97/23/EC (Note 1)
Note : See table 2 for detail.
Approved hazardous location certifications:
Model: LF414/LF600 and LF414/LF602
cFMus explosion proof:
FM Class I, Division 2, Groups A,B,C, and D. FM Class II, Division 2, Groups E, F and G. FM Class III.
Detector and converter combination:
LF410/LF600: Combined type for standard specification.
LF410/LF602: Separate type for standard specification.
LF414/LF600: Combined type with Ex approval of Class I, Division 2 (cFMus).
LF414/LF602: Separate type with Ex approval of Class I, Division 2 (cFMus).
Model LF410 and LF414 Detectors
Mounting style: Wafer type
Fluid pressure:
Note: The test pressure before shipping from the factory is equal to twice the nominal pressure rating of the customer specified flange connection during 15 minutes.
Connection flange standards: ANSI 150, ANSI 300, BS10 and 16, DIN PN10 and PN16, JIS10K, JIS16K and JIS20K
Principal materials:
Case — 25 to 100mm (1” to 4”): stainless steel 15, 150, and 200mm (1/2”, 6”, and 8”): carbon steel
Linings — 15 to 100mm (1/2” to 4”): Ceramic tube (std.) & teflon PFA (opt.)
150 and 200mm (6” to 8”): teflon PFA Electrodes — 316L stainless steel (std.) Grounding rings — 316 stainless steel (std.)
Note: See Table 3 for optional materials and other related information.
Measuring tube material — 304 stainless steel (in case of Teflon PFA lining)
Coating — 25 to 100mm (1” to 4”): no coating (stainless steel body).
15, 150, and 200mm (1/2”, 6”, and 8”): phthalic acid resin coating with
Heat shock resistance : for a ceramic tube detector Heating: ΔT ≤ 150 °C/0.5sec (302°F/0.5sec) Cooling: ΔT ≤ 100 °C/0.5sec (212°F/0.5sec)
Note: The above means that the ceramic tube detector
withstands the shock of sudden heating (temperature difference 150°C or less per 0.5seconds) and sudden cooling (temperature difference 100°C or less per 0.5seconds).
Dimensions and weights: See Figures 3 to 8.
Cable connection port: for separate type detectors. Cable glands —
LF410: without cFMus approval Provided as standard R(PT) 1/2 male screws.
LF414: with cFMus approval
Not provided
Applicable diameter — 11 to 13mm
(0.433 to 0.512 inch)
Model LF600 and LF602 converters
Input signals
Analog signal — the voltage signal from detector, proportional to process flow rate (for LF602 separate type converter).
Digital input DI (opt.)
Signal type: 20 to 30Vdc voltage signal
Input resistance: 2.7kΩ
Number of inputs: one point
DI function — One of the following functions can be assigned to the optional DI signal.
Range switching — Selects either the higher or lower range in the unidirectional or bidirectional
Totalizer control — Starts and stops the
2