LF410/LF600 LF410/LF602
totalizer.
Zero adjustment — Executes zero adjustment
Output signals
Current output:
Note: The current output cannot be used with the
Digital outputs — One point (std.) and one more point is optionally available as follows.
Digital output DO1 (std.):
Output type: Transistor open collector
Number of outputs: One point
Output capacity: 30Vdc, 200mA maximum
Digital output DO2 (opt.):
Output type: Solidstate relay output (non polarity)
Number of outputs: One point
Output capacity: 150Vdc, 150mA maximum or 150 V ac (peak to peak), 100mA maximum
DO1 and DO2 functions — One of the following functions can be assigned to DO1 (std.) and/or DO2 (opt.)
•Pulse output (available only for DO1,DO2) Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)
3.6 to 360,000 pulses/hr (DO2) (Over 3,600,000 pulses/hr,
Pulse width: 0.5 to 500ms (but less than half of the period for 100% flow rate)
Note: The same and simultaneous pulse is not available between DO1 and DO2.)
•
•High, High high, Low, and/or Low low alarm outputs (Note 2)
•Empty pipe alarm output
•Digital Output Active Status (DO1 and DO2) (Note 2)
•Preset count output
•Converter failure alarm output
Note 1: Two outputs (DO1 and DO2) are needed for
Note 2: Normal Open (default set) or Normal Close
is selected for alarm outputs when programming.
The Status when power failure is kept to Normal Open.
Communications output :
•HART(std.)
Digital signal is superimposed on
•Conforms to HART protocol Load resistance: 240 to 750Ω Load capacitance: 0.25µF maximum Load inductance: 4mH maximum
•PROFIBUS (opt.) Protocol :
Consumption electric current of bus:less than 16mA
Manufacture
Standard
Slave address :
Profile : Profile Ver.3.01 for Process Control Devices
Function blocks : AI(Flow)×1 , Totalizer×1
LCD display: Full
The data on the LCD inside the converter can rotate to 90, 180, and 270 degrees by a software, without rotating the indicator itself. (Combined type only)
Parameter settings — Parameters can be set as follows:
•IR Switches: Three key switches are provided to set configuration parameters.
•Digital communication: The AF900
parameters.
•Zero adjustment: Zero point adjustment can be started by pressing the switch in the converter.
Damping: 0.5 to 60 seconds (selectable in 1 second increments)
Zero and span calibration:
Conditions when power fails: The outputs and display will remain as follows when power fails. Parameter setting values are stored in
•Current output: 0mAdc
•Digital output: OFF
•LCD display: No display
•PROFIBUS: No communication
Power supply: One of the following can be selected:
•100 to 240Vac, 50/60Hz (std.) (allowable voltage 80 to 264Vac)
•24Vdc (allowable voltage 18 to 36Vdc)
•110Vdc (allowable voltage 90 to 130Vdc)
Surge protection: Arresters are installed in the power supply, and current signal output circuit.
Case: Aluminum alloy (equivalent of IP 67)
Coating: Acrylic
3