(3) Connecting the input signal cable

6F8A0870

(3) Connecting the input signal cable

Separate

Strip the sheath from the end of each conductor of a 2-core individually shielded cable as shown in Figure

5.4.Twist those shields and cover them with a thermal contraction tube or vinyl tube not to make contact with the case or core wires. Then attach an M3.5 crimping terminal with insulated sleeve as shown in Figure

5.3.Connect a crimping terminal to the A and B terminals on the terminal block and connect to each G terminal of the detector and converter.

 

M3.5 crimping terminal

A Black

 

Thermal contraction tube or vinyl tube

 

 

GShield B White

25~45mm

Figure 5.3 Terminal Treatment of Flow Rate Signal Cable

Notes on signal cable shield processing work

When stripping an external sheath, intermediate and insulated sheath, be careful not to scratch or cut the internal conductors and shield mesh. Do not disjoint the shield mesh but treat it as shown in Figure 5.9.

a. Open the shield mesh with a pencil or the like.

Coated wire

b.Pull out the internal coated wires from the hole of the shielded mesh.

Shield mesh

c.Pull out all internal coated wires and extend the shield mesh wire.

Figure 5.4 Treating the Signal Cable Shield Mesh