5.4.2 Cable Connection

6F8A0870

5.4.2 Cable Connection

Connect and install the terminal-treated cables to the terminal block.

*Connect the cables to the terminal block securely. A loose connection may cause incorrect measurement. After connecting a cable, try to pull it to check whether it has been connected securely.

Referring to combined converter's manuals of "Connections and Grounding", connect each cable to the terminal block. Tighten the screws of the terminal block tightly to ensure a secure connection. A loose connection may cause incorrect measurement. After connecting a cable, try to pull it to see whether it has been connected securely.

Phillips screwdriver

* The appropriate torque for tightening the terminal board screw is 1.2 N.m.

Terminal block

Figure 5.5 Connecting a Cable to Terminal Block

- 40 -

Page 41
Image 41
Toshiba LF414 manual Cable Connection, - 40 -, 6F8A0870, 5 Connecting a Cable to Terminal Block