Toshiba LF414 Wiring 5.4.1 Terminal Treatment of Cables, Separate 2 Excitation cable, - 38 -

Models: LF414

1 65
Download 65 pages 63.14 Kb
Page 39
Image 39
5.4Wiring
M3.5 crimping terminal

6F8A0870

5.4Wiring

5.4.1Terminal Treatment of Cables

Follow the procedures below to treat the terminals (at the converter side) of various cables and install the cables to the terminal block. Use appropriate cables based on the description in Section 5.1 "Cables." Crimp a round type insulated crimp-type terminal to the end of the cables.

(1) Power cable, current output cable, digital I/O cables

The necessary cables should be ordered from the person responsible for the installation. Strip the sheath of each conductor as shown in Figure 5.1 and attach a crimping terminal with insulated sleeve to it. The size of the crimping terminal is as follows:

M4: Integral type LF600F and LF610F

M3.5: Integral type LF620F, Separate type LF602F, LF612F and LF622F

Connect the power cable to terminal blocks L1 and L2.

Connect the current output cable to terminal blocks + and -.

Connect the digital I/O cable to terminal blocks D1, D01, D02 and COM, as required.

Crimping terminal

M4: LF600F and LF610F

M3.5: LF602F, LF612F, LF620F and LF622F

 

25~45mm

Figure 5.1 Terminal Treatment of Power Cable, Current Output Cable and Digital I/O cable

Separate (2) Excitation cable

Strip the sheath from the end of each conductor as shown in Figure 5.2, attach an M3.5 crimping terminal with insulated sleeve, and connect it to the terminal blocks X and Y. Connect the red conductor to terminal block E.

X Black

ERed

Y White

25~45mm

Figure 5.2 Terminal Treatment of Excitation Cable

- 38 -

Page 39
Image 39
Toshiba LF414 Wiring 5.4.1 Terminal Treatment of Cables, Power cable, current output cable, digital I/O cables, - 38 -