Toshiba LF414 Flow rate indicated is not correct, - 45 -, 6F8A0870, Start, set?, as specified?

Models: LF414

1 65
Download 65 pages 63.14 Kb
Page 46
Image 46
7.2.2Flow rate indicated is not correct

6F8A0870

7.2.2Flow rate indicated is not correct

 START START set?

Is the flow range correctly

NO

 

set?

 

YES

 

Is zero point correctly set?

NO

 

YES

 

Is the excitation current

NO

value as stated on the

 

flow direction tag?

 

YES

 

Is the inside wall of detector

YES

pipe contaminated?

 

NO

 

Is the output load resistance

NO

as specified?

 

YES

 

Are there two load

YES

resistors connected to

 

the output in parallel?

 

NO

 

Is accuracy calculated as follows? NO

(Measured flow rate)-(Actual flow rate)

×100%

Actual flow rate

 

YES

Contact your nearest Toshiba representative.

Set correctly.

Refer to combined converter's manual.

Perform the zero adjustment.

Refer to combined converter's manual.

Set correctly.

Refer to combined converter's manual.

Clean the inside of the detector pipe.

Design the output circuit so that the output load resistance is less than

750Ω.

Connect the two load resistors in series, if necessary.
Calculate as shown on the left.

- 45 -

Page 46
Image 46
Toshiba LF414 Flow rate indicated is not correct, - 45 -, 6F8A0870, Start, set?, Is the excitation current, as specified?