•Key M : MODE

Model 5100-02-IT Combustible Gas Sensor Module

7.10APPENDIX J: Menu Key for 5100-02-IT Gas Sensor Module

Key M : MODE

Key E : ENTER

Key UP (+)

Key DOWN (-)

MAIN MENU

Use the M key to access each of the 4 main sections of the menu.

MM

MMM

 

 

 

 

 

 

 

 

 

-

ALMRSET

:

-

CALIB

:

-

SETUP

:

-

MAINT

:

-

EXIT?

ALARM RESET MENU

ALMRSET:

E– RESET (Resets current alarm)

CALIB MENU

CALIB:

ECAL-0% (Calibrate Zero)

E – Ack (Acknowledges Zero)

*50%-SPAN (Last span gas used – initially set at 50%)

 

 

 

 

 

25%-SPAN

50%-SPAN (Select Span gas value from below)

75%-SPAN

Adj-SPAN (To select an alternate span gas value)

 

 

 

 

 

 

E

50%-SPAN

 

 

 

 

 

 

 

51%-SPAN (Change span value to desired, once completed E starts calibration)

 

 

 

 

 

 

EC XX%LEL (Reading the value of the calibration gas being applied)

- CAL-OK (if successful) or CAL-FAIL (if unsuccessful)