Univex 5100-02-IT Gas Factor Example, 4-20mA Adjustment Example, RS-485Adjustment Example, Page

Models: 5100-02-IT IT Series

1 41
Download 41 pages 37.73 Kb
Page 21
Image 21
Gas Factor Example

Model 5100-02-IT Combustible Gas Sensor Module

Gas Factor Example

M

E

Enter

GasFactr

S.P. Function - Gas Factor Adjust

*C

S

T

M

E

Enter

Factr100

Select [E] to select or S or T to adjust factor

 

 

 

 

S

T

number and press [E]

 

M

E

Enter

ACK

Acknowledgement of new Gas Factor Value

 

S

T

 

 

 

 

 

 

 

 

 

 

 

4-20 mA Adjustment Example

 

 

 

 

 

 

 

M

E

Enter

Calib

S.P. Function - Calib Adjust

*D

S T

M

E

Enter

Out: 4mA

Use S or T keys to select 4 mA or 20mA

 

S

T

 

M

E

Enter

4mA

Selects 4 mA

 

S

T

 

M

E

Enter

ACK

Momentary Acknowledge of new Set Point

 

S

T

 

M

E

Mode

Calib

S.P. Function - Calib Adjust

 

S

T

 

M

E

Down

CalibOut

S.P. Function - Output during Calibration Adjust

 

S

T

 

M

E

Enter

Track

Use S or T keys to select Track, Zero, C1.50mA, C4.00mA

S

T

 

 

 

 

Track = Output during calibration tracks the calibration gas,

 

 

 

 

Zero = Output is Zero during calibration,

 

 

 

 

 

C1.50mA = Output during calibration is 1.50 mA, C4.00mA =

 

 

 

 

Output during calibration is 4.0 mA

 

M

E

Enter

*Track

* = Current selection

 

S

T

 

M

E

Mode

Calib

S.P. Function - Calib Adjust

 

S T

 

M

E

Down

Tbl Out

S.P. Function - output during Trouble Alarm

 

S T

 

M

E

Enter

*T2.50mA

Current Valve 2.5mA for Trouble

 

S T

 

M

E

Down

User mA

S.P. Function - User Selectable mA Valve

 

S T

 

M

E

Enter

*T2.50mA

Use S or T keys to select mA Valve between 0.5

 

S T

and 0.7mA

 

M

E

Enter

*T1.75mA

Selctd 1.75mA for volume during trouble

 

S T

 

 

 

 

 

 

 

 

 

 

 

RS-485 Adjustment Example

 

 

 

 

 

 

 

M

E

Enter

Address

S.P. Function - RS-485 Address Adjust

*E

S

T

M

E

Enter

Addr:016

Use S or T keys to enter new address

 

S T

 

 

 

 

 

 

 

M

E

Enter

ACK

New address selected

 

S

T

 

M

E

Enter

Address

S.P. Function - RS-485 Address Adjust

 

S

T

 

M

E

Down

Baud

S.P. Function - RS-485 Baud Rate Adjust

 

S

T

 

M

E

Enter

*38400

Press [E] to select or [S] or [T] to select another

 

S T

 

Table 4-3B

Set-Up Configuration

Page: 19

Page 21
Image 21
Univex 5100-02-IT Gas Factor Example, 4-20mA Adjustment Example, RS-485Adjustment Example, 3B Set-UpConfiguration, Page