Emerson 5081-T Using Hold, Entering the Security Code, Retrieving a Lost Security Code

Page 51

MODEL 5081-T

SECTION 5.0

 

DISPLAY AND OPERATION

PROGRAM

5G

EXIT

"""

ENTER

5.8 SECURITY

5.8.1General. Use the programmable security code to protect program and calibration settings from accidentally being changed. The transmitter is shipped with the security fea- ture disabled.

5.8.2Entering the Security Code.

1. If calibration and program settings are protected with a security code, pressing PROG or CAL on the infrared remote controller causes the Id screen to appear.

2.Use the editing keys to enter the security code. Press ENTER .

3.If the security code is correct, the first sub-menu appears. If the security code is incor- rect, the process display reappears.

5.8.3 Retrieving a Lost Security Code.

1.If the security code has been forgotten, enter 555 at the Id prompt and press ENTER . The transmitter will display the present code.

2.Press EXIT to return to the process display.

3.Press PROG or CAL . The Id screen appears.

4.Use the editing keys to enter the security code just shown; then press ENTER .

5.The first sub-menu under the selected menu will appear.

5.9 USING HOLD

During calibration, the sensor may be exposed to solutions having concentration outside the normal range of the process. To prevent false alarms and undesired operation of chemical dosing pumps, place the transmitter in hold during calibra- tion. Activating hold keeps the transmitter output at the last value or sends the output to a previously determined value. See Section 7.2, Output Ranging, for details.

After calibration, reinstall the sensor in the process stream. Wait until readings have stabilized before deactivating Hold.

To activate or deactivate Hold:

1.Press HOLD on the remote controller.

2.The HoLd prompt appears in the display. Press ￿ or ￿ to toggle Hold between On and OFF.

3.Press ENTER to save.

45

Image 51
Contents Model 5081-T Rev. Level Date Essential InstructionsModel 5081-T Table of Contents CONT’D List of Figures Number TitleIii List of Tables List of Figures CONT’DSimple Menu Structure Section Description and SpecificationsHart Recommended Sensors AtexRecommended Ranges for Toroidal Sensors Transmitter Display During Calibration and Programming I T N E X THart Interface Devices Overview of Hart CommunicationFoundation Fieldbus AMS Main Menu Tools Orienting the Display Board Section InstallationMechanical Installation Unpacking and InspectionMounting on a Flat Surface Pipe Mounting 40308103Recommended Sensors Inductive LoopsSensor Selection Model 5081-T Sensor SelectionModel 5081-T-HT Power SUPPLY/CURRENT LoopPower Supply and Load Requirements. Refer to Figure Power Supply-Current Loop WiringTypical Fieldbus Network Electrical Wiring Configuration Power Supply Wiring for Model 5081-T-FF/FIWiring Through a Junction BOX Section WiringSensor Wiring Electrical Installation Sensor WiringPower Wiring Model 5081-T-FF Power Wiring Model 5081-T-HTWiring Model 242 sensor to Model 5081-T transmitter Electrical InstallationSection Section Section Intrinsically Safe & Explosion Proof IRC Infrared Remote ControlFM Explosion-Proof Installation for Model 5081-T-HT FM Intrinsically Safe Installation for Model 5081-T-HT CSA Intrinsically Safe Installation for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT FM Explosion-Proof Installation for Model 5081-T-FF FM Intrinsically Safe Installation for Model 5081-T-FF 10. CSA Intrinsically Safe Installation for Model 5081-T-FF 11. Atex Intrisically Safe Label for Model 5081-T-FF 12. Atex Intrinsically Safe Installation for Model 5081-T-FF FI-T 9241515-01 15. FM Intrinsically Safe Installation for Model 5081-T-FI 9241516-01 17. CSA Intrinsically Safe Installation for Model 5081-T-FI 18. Atex Intrisically Safe Label for Model 5081-T-FI 19. Atex Intrinsically Safe Installation for Model 5081-T-FI I T N E X T Section Display and OperationDisplays S UInfrared Remote Controller Infrared Remote Controller IRC KEY FunctionsQuick Start for Model 5081-T-HT Hart To reset transmiter to factory default settingsQuick Start for Model 5081-T-FF/FI Foundation Fieldbus Calibration Program Menu Tree Conductivity25.0C 12.00mA Process Display Program Menu Mnemonics Default Settings Diagnostic MessagesDefault Settings for Model 5081-T-HT Retrieving a Lost Security Code Using HoldSecurity Entering the Security CodeSection START-UP and Calibration Accessing the Calibrate MenuModel 5081-T-FF Calibrate MenuOutput Cal Procedure9URU Temp SlopeU91 5ES-U1U17!91 GM!91Section Programing GeneralModel 5081-T-HT Model 5081-T-FF Output 5081-HT only PurposeDefinitions Uhtu Procedure9VUVU 4QGProcess temperature is. Press Enter to save TempU7-P H2SO4L DisplayG5-A HCLGgshtt Hart8VO!UT Setup CustPVO!RUT!!% SHIG12-U DefaultSetting Measurement Range Over Range Warning Limit RangePID Section Foundation Fieldbus OperationSection Operation with Model Connecting the CommunicatorMaking Hart related settings from the keypad OperationOff-line and On-line Operation C/T-HT HART/Model 375 Menu Tree 1 PV LRV PV URVC/T-HT HART/Model 375 Menu Tree 2 LSL USLAO1 LRV URVC/T-HT HART/Model 375 Menu Tree 4 PV AO PV LRV PV URVMaxnotify Resource FaultstateBlockerr Featuresel Rsstate Faultstate LimnotifyTagdesc AlarmhysStrev BlockerrStrev Statusopts Strategy Alertkey HihilimPvftime LowcutIoopts LolimLolopri LololimBkcalhys Ffgain ResetRate Reset Sphilim SplolimBypass BaltimeAlertkey Outhilim ControloptsSphilim Strev Rate TagdescDvlopri DvhipriC/T-FF/FI Model 375 Menu Tree 9 C/T-FF/FI Model 375 Menu Tree 10 Section Diagnosis and Troubleshooting OverviewCurrent Operating Menu Idvnut DLSTqiu 4DSGFault Conditions Disabling FaultsNON- Disabling Warnings 9WHS!S-PJ1 Hssqs9IITHU!1SS Field Troubleshooting Quick Troubleshooting GuideQuick Troubleshooting Guide Symptom ActionTroubleshooting Flow Chart Systematic TroubleshootingRTD Resistance Values RTD Resistance ValuesFormula ExampleWhen a Fault exists When a Warning condition existsCPU Fail Troubleshooting When a Fault or Warning Message is ShowingTemp HI Temp LORtd FAIL, Temp HI, and Temp LO CPU Fail and ROM Fail 0 C 2 1 0 % mA Section MaintenancePreventative Maintenance Transmitter MaintenanceExploded View of Model 5081-T Transmitter Replacement Parts for Model 5081-T TransmitterOutput Logic Section Theory of OperationConductivity Hart CommunicationGeneral Warranty RepairNON-WARRANTY Repair Section Return of MaterialThis page intentionally left blank Return of Material WarrantySpecifications subject to change without notice