Emerson 5081-T Off-line and On-line Operation, Making Hart related settings from the keypad

Page 66

MODEL 5081-T

SECTION 9.0

 

OPERATION WITH MODEL 375

9.3Operation

9.3.1Off-line and On-line Operation

The Model 375 Communicator features off-line and on-line communications. On-line means the communicator is connected to the transmitter in the usual fashion. While the communicator is on line, the operator can view meas- urement data, change program settings, and read diagnostic messages. Off-line means the communicator is not connected to the transmitter. When the communicator is off line, the operator can still program settings into the communicator. Later, after the communicator has been connected to a transmitter, the operator can transfer the programmed settings to the transmitter. Off-line operation permits settings common to several transmitters to be easily stored in all of them.

9.3.2Making HART related settings from the keypad

Calibrate

Hold

Program

Display

 

 

 

 

Output

Temp

Measurement

>>

 

 

 

 

Security

HART

 

>>

 

 

 

 

DevID

PollAddrs

Burst

Preamble

 

 

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose >>.

3.Choose HART.

4.To display the device ID, choose DevID. To change the polling address, choose PollAddrs. To make burst mode settings, choose Burst. To change the preamble count, choose Preamble.

9.3.3Menu Tree

The menu tree for the Model 375 HART communicator is on the following page. The menu tree for the Model 375 FOUNDATION Fieldbus communicator immediately follows.

Note: Model 375 Communicator fully supports FOUNDATION Fieldbus on Model 5081-T-FF.

60

Image 66
Contents Model 5081-T Essential Instructions Rev. Level DateModel 5081-T Table of Contents CONT’D List of Figures Number TitleIii List of Figures CONT’D List of TablesSection Description and Specifications Simple Menu StructureHart Recommended Sensors AtexRecommended Ranges for Toroidal Sensors I T N E X T Transmitter Display During Calibration and ProgrammingOverview of Hart Communication Hart Interface DevicesFoundation Fieldbus AMS Main Menu Tools Unpacking and Inspection Section InstallationMechanical Installation Orienting the Display BoardMounting on a Flat Surface 40308103 Pipe MountingModel 5081-T Sensor Selection Inductive LoopsSensor Selection Recommended SensorsPower Supply-Current Loop Wiring Power SUPPLY/CURRENT LoopPower Supply and Load Requirements. Refer to Figure Model 5081-T-HTPower Supply Wiring for Model 5081-T-FF/FI Typical Fieldbus Network Electrical Wiring ConfigurationSensor Wiring Section WiringSensor Wiring Electrical Installation Wiring Through a Junction BOXPower Wiring Model 5081-T-HT Power Wiring Model 5081-T-FFElectrical Installation Wiring Model 242 sensor to Model 5081-T transmitterSection Section IRC Infrared Remote Control Section Intrinsically Safe & Explosion ProofFM Explosion-Proof Installation for Model 5081-T-HT FM Intrinsically Safe Installation for Model 5081-T-HT CSA Intrinsically Safe Installation for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT FM Explosion-Proof Installation for Model 5081-T-FF FM Intrinsically Safe Installation for Model 5081-T-FF 10. CSA Intrinsically Safe Installation for Model 5081-T-FF 11. Atex Intrisically Safe Label for Model 5081-T-FF 12. Atex Intrinsically Safe Installation for Model 5081-T-FF FI-T 9241515-01 15. FM Intrinsically Safe Installation for Model 5081-T-FI 9241516-01 17. CSA Intrinsically Safe Installation for Model 5081-T-FI 18. Atex Intrisically Safe Label for Model 5081-T-FI 19. Atex Intrinsically Safe Installation for Model 5081-T-FI S U Section Display and OperationDisplays I T N E X TInfrared Remote Controller IRC KEY Functions Infrared Remote ControllerTo reset transmiter to factory default settings Quick Start for Model 5081-T-HT HartQuick Start for Model 5081-T-FF/FI Foundation Fieldbus Process Display Menu Tree Conductivity25.0C 12.00mA Calibration ProgramProgram Menu Mnemonics Diagnostic Messages Default SettingsDefault Settings for Model 5081-T-HT Entering the Security Code Using HoldSecurity Retrieving a Lost Security CodeAccessing the Calibrate Menu Section START-UP and CalibrationCalibrate Menu Model 5081-T-FFTemp Slope Procedure9URU Output CalGM!91 5ES-U1U17!91 U91Section Programing GeneralModel 5081-T-HT Model 5081-T-FF Output 5081-HT only PurposeDefinitions 4QG Procedure9VUVU UhtuProcess temperature is. Press Enter to save TempU7-P HCL DisplayG5-A H2SO4LHart GgshttSHI Setup CustPVO!RUT!!% 8VO!UTRange DefaultSetting Measurement Range Over Range Warning Limit G12-USection Foundation Fieldbus Operation PIDConnecting the Communicator Section Operation with ModelMaking Hart related settings from the keypad OperationOff-line and On-line Operation PV LRV PV URV C/T-HT HART/Model 375 Menu Tree 1LSL USL C/T-HT HART/Model 375 Menu Tree 2LRV URV AO1PV AO PV LRV PV URV C/T-HT HART/Model 375 Menu Tree 4Limnotify Resource FaultstateBlockerr Featuresel Rsstate Faultstate MaxnotifyBlockerr AlarmhysStrev TagdescLowcut Alertkey HihilimPvftime Strev Statusopts StrategyLololim LolimLolopri IooptsSphilim Splolim ResetRate Reset Bkcalhys FfgainBaltime BypassStrev Rate Tagdesc ControloptsSphilim Alertkey OuthilimDvhipri DvlopriC/T-FF/FI Model 375 Menu Tree 9 C/T-FF/FI Model 375 Menu Tree 10 Section Diagnosis and Troubleshooting OverviewCurrent Operating Menu 4DSG DLSTqiu IdvnutFault Conditions Disabling FaultsNON- Disabling Warnings 9WHS!S-PJ1 Hssqs9IITHU!1SS Symptom Action Quick Troubleshooting GuideQuick Troubleshooting Guide Field TroubleshootingSystematic Troubleshooting Troubleshooting Flow ChartExample RTD Resistance ValuesFormula RTD Resistance ValuesWhen a Warning condition exists When a Fault existsTemp LO Troubleshooting When a Fault or Warning Message is ShowingTemp HI CPU FailRtd FAIL, Temp HI, and Temp LO CPU Fail and ROM Fail Transmitter Maintenance Section MaintenancePreventative Maintenance 0 C 2 1 0 % mAReplacement Parts for Model 5081-T Transmitter Exploded View of Model 5081-T TransmitterHart Communication Section Theory of OperationConductivity Output LogicSection Return of Material Warranty RepairNON-WARRANTY Repair GeneralThis page intentionally left blank Warranty Return of MaterialSpecifications subject to change without notice