MODEL | SECTION 7.0 |
| PROGRAMMING |
7.4 DISPLAY
7.4.1 Purpose
This section describes how to do the following:
1.Configure the transmitter to measure conductivity, resistivity, or set up a custom curve
2.Set the temperature units to °C or °F
3.Set the output to current or percent of full scale
4.Enter a security code.
7.4.2 Definitions
1.MEASUREMENT. The transmitter can be configured to measure conductivity in mS/cm or resistivity in MegOhms, or configured with a
2.OUTPUT CURRENT. The transmitter generates a
5. SECURITY CODE. The security code unlocks the transmitter and allows access to all menus.
7.4.3 Procedure
PROGRAM
G5<:[-A
EXIT NEXT ENTER
PROGRAM
UA:1 /QPGVF
EXITENTER
1.Press PROG on the remote controller.
2.Press NEXT until the diSPLAy submenu appears. Press ENTER.
3.Press or to display the desired measurement. Press ENTER to save.
| COnduc | Conductivity |
|
| nAOH | Sodium Hydroxide |
|
| HCL | Hydrochloric Acid |
|
| H2SO4L | Sulfuric Acid |
|
| H2SO4H | Sulfuric Acid |
|
| CuSt | Custom Curve |
|
|
|
|
|
4.tEMP will appear. Press or to display the desired temperature reading, C or F. Press ENTER.
5.OUtPUt will appear. Press or to display the desired
6.COdE will appear. Using the arrow keys on the remote control, enter the desired
54