Emerson 5081-T Recommended Sensors, Atex, Recommended Ranges for Toroidal Sensors

Page 9

MODEL 5081-T

SECTION 1.0

 

DESCRIPTION AND SPECIFICATIONS

1.2.3TRANSMITTER SPECIFICATIONS @ 25°C Measured Range*: 50 to 2,000,000 µS/cm (see chart) Accuracy: ± 1.0% of reading

Repeatability: ± 0.25% of reading

Stability: 0.25% of output range/month, non-cumulative

Ambient Temperature Coefficient: ± 0.2% of FS/°C

Temperature Slope Adjustment: 0-5%/° C

%Concentration Ranges: Sodium Hydroxide: 0 to 15% Hydrochloric Acid: 0 to 16%

Sulfuric Acid: 0 to 25% and 96 to 99.7%

1.2.4 LOOP SPECIFICATIONS

Loop Accuracy: With a standard Model 228 or 225 sen- sor with 20' cable, laboratory accuracy at 25°C can be as good as ±2% of reading and ±50 µS/cm.

To achieve optimum performance, standardize the sen- sor in the process at the conductivity and temperature of interest.

Results under real process conditions, at different tem- peratures, or using other sensors may differ from above.

RTD accuracy: Utilizing a perfect 100 Ohm RTD after 1 point temperature standardization, temperature reading can be as good as ±0.5°C.

1.3 HAZARDOUS LOCATION APPROVAL

Intrinsic Safety:

Class I, II, III, Div. 1

Groups A-G

T4 Tamb = 70°C

Exia Entity

Class I, Groups A-D

Class II, Groups E-G

Class III

T4 Tamb = 70°C

ATEX

1180

II1 G Baseefa03ATEX0399 EEx ia IIC T4 Tamb = -20°C to +65°C

Non-Incendive:

Class I, Div. 2, Groups A-D

Dust Ignition Proof

Class II & III, Div. 1, Groups E-G

NEMA 4X Enclosure

RECOMMENDED SENSORS:

Model 222 Flow-Through

Model 225 Clean-In-Place (CIP)

Model 226 Submersion/Insertion

Model 228 Submersion/Insertion/Retractable

Model 242* Flow-Through

*no I.S. approval for loops of 5081-T with 242-06 or 242-08

Class I, Div. 2, Groups A-D

Suitable for

Class II, Div. 2, Groups E-G

T4 Tamb = 70°C

Explosion-Proof:

Class I, Div. 1, Groups B-D

Class II, Div. 1, Groups E-G

Class III, Div. 1

Class I, Groups B-D

Class II, Groups E-G

Class III

Tamb = 65°C max

RECOMMENDED RANGES FOR TOROIDAL SENSORS

 

 

Conductivity Sensor

 

 

 

 

 

 

 

Model Number

226

228

225

222 (1in.)

222 (2 in.)

242

 

Nominal Cell Constant

1.0

3.0

3.0

6.0

4.0

*

 

Minimum Conductivity (S/cm)

50

200

200

500

500

100*

 

 

 

 

 

 

 

 

 

Maximum Conductivity (S/cm)

1,000,000

2,000,000

2,000,000

2,000,000

2,000,000

2,000,000*

 

 

 

 

 

 

 

 

 

* Model 242 values depend on sensor configuration and wiring.

 

 

 

3

Image 9
Contents Model 5081-T Rev. Level Date Essential InstructionsModel 5081-T Table of Contents CONT’D List of Figures Number TitleIii List of Tables List of Figures CONT’DSimple Menu Structure Section Description and SpecificationsHart Recommended Sensors AtexRecommended Ranges for Toroidal Sensors Transmitter Display During Calibration and Programming I T N E X THart Interface Devices Overview of Hart CommunicationFoundation Fieldbus AMS Main Menu Tools Mechanical Installation Section InstallationUnpacking and Inspection Orienting the Display BoardMounting on a Flat Surface Pipe Mounting 40308103Sensor Selection Inductive LoopsModel 5081-T Sensor Selection Recommended SensorsPower Supply and Load Requirements. Refer to Figure Power SUPPLY/CURRENT LoopPower Supply-Current Loop Wiring Model 5081-T-HTTypical Fieldbus Network Electrical Wiring Configuration Power Supply Wiring for Model 5081-T-FF/FISensor Wiring Electrical Installation Section WiringSensor Wiring Wiring Through a Junction BOXPower Wiring Model 5081-T-FF Power Wiring Model 5081-T-HTWiring Model 242 sensor to Model 5081-T transmitter Electrical InstallationSection Section Section Intrinsically Safe & Explosion Proof IRC Infrared Remote ControlFM Explosion-Proof Installation for Model 5081-T-HT FM Intrinsically Safe Installation for Model 5081-T-HT CSA Intrinsically Safe Installation for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT FM Explosion-Proof Installation for Model 5081-T-FF FM Intrinsically Safe Installation for Model 5081-T-FF 10. CSA Intrinsically Safe Installation for Model 5081-T-FF 11. Atex Intrisically Safe Label for Model 5081-T-FF 12. Atex Intrinsically Safe Installation for Model 5081-T-FF FI-T 9241515-01 15. FM Intrinsically Safe Installation for Model 5081-T-FI 9241516-01 17. CSA Intrinsically Safe Installation for Model 5081-T-FI 18. Atex Intrisically Safe Label for Model 5081-T-FI 19. Atex Intrinsically Safe Installation for Model 5081-T-FI Displays Section Display and OperationS U I T N E X TInfrared Remote Controller Infrared Remote Controller IRC KEY FunctionsQuick Start for Model 5081-T-HT Hart To reset transmiter to factory default settingsQuick Start for Model 5081-T-FF/FI Foundation Fieldbus 25.0C 12.00mA Menu Tree ConductivityProcess Display Calibration ProgramProgram Menu Mnemonics Default Settings Diagnostic MessagesDefault Settings for Model 5081-T-HT Security Using HoldEntering the Security Code Retrieving a Lost Security CodeSection START-UP and Calibration Accessing the Calibrate MenuModel 5081-T-FF Calibrate Menu9URU ProcedureTemp Slope Output CalU17!91 5ES-U1GM!91 U91Section Programing GeneralModel 5081-T-HT Model 5081-T-FF Output 5081-HT only PurposeDefinitions 9VUVU Procedure4QG UhtuProcess temperature is. Press Enter to save TempU7-P G5-A DisplayHCL H2SO4LGgshtt HartPVO!RUT!!% Setup CustSHI 8VO!UTSetting Measurement Range Over Range Warning Limit DefaultRange G12-UPID Section Foundation Fieldbus OperationSection Operation with Model Connecting the CommunicatorMaking Hart related settings from the keypad OperationOff-line and On-line Operation C/T-HT HART/Model 375 Menu Tree 1 PV LRV PV URVC/T-HT HART/Model 375 Menu Tree 2 LSL USLAO1 LRV URVC/T-HT HART/Model 375 Menu Tree 4 PV AO PV LRV PV URVBlockerr Featuresel Rsstate Faultstate Resource FaultstateLimnotify MaxnotifyStrev AlarmhysBlockerr TagdescPvftime Alertkey HihilimLowcut Strev Statusopts StrategyLolopri LolimLololim IooptsRate Reset ResetSphilim Splolim Bkcalhys FfgainBypass BaltimeSphilim ControloptsStrev Rate Tagdesc Alertkey OuthilimDvlopri DvhipriC/T-FF/FI Model 375 Menu Tree 9 C/T-FF/FI Model 375 Menu Tree 10 Section Diagnosis and Troubleshooting OverviewCurrent Operating Menu Tqiu DLS4DSG IdvnutFault Conditions Disabling FaultsNON- Disabling Warnings 9WHS!S-PJ1 Hssqs9IITHU!1SS Quick Troubleshooting Guide Quick Troubleshooting GuideSymptom Action Field TroubleshootingTroubleshooting Flow Chart Systematic TroubleshootingFormula RTD Resistance ValuesExample RTD Resistance ValuesWhen a Fault exists When a Warning condition existsTemp HI Troubleshooting When a Fault or Warning Message is ShowingTemp LO CPU FailRtd FAIL, Temp HI, and Temp LO CPU Fail and ROM Fail Preventative Maintenance Section MaintenanceTransmitter Maintenance 0 C 2 1 0 % mAExploded View of Model 5081-T Transmitter Replacement Parts for Model 5081-T TransmitterConductivity Section Theory of OperationHart Communication Output LogicNON-WARRANTY Repair Warranty RepairSection Return of Material GeneralThis page intentionally left blank Return of Material WarrantySpecifications subject to change without notice