Emerson 5081-T instruction manual 5ES-U1, U17!91, GM!91, U91, 9PU, 1P9S, U17!-G6

Page 55

MODEL 5081-T

SECTION 6.0

 

START-UP AND CALIBRATION

 

TABLE 6-1. CALIBRATE MENU MNEMONICS

 

 

/-[5ES-U1

Calibrate menu header

/-[

Sensor calibration

U17:!<[9:1

Sub-menu header

-GM!<[9:1

Sub-menu header

U<[9:1

Slope adjustment %/°C

/1[[!/9P<U

Sub-menu header

<1P<9S!

Sub-menu header

<1P<9S!"

Sensor "0" (performed in air)

U17:!-G6

Sub-menu header

U17:

Temperature adjustment °C/°F

 

 

49

Image 55
Contents Model 5081-T Rev. Level Date Essential InstructionsModel 5081-T Table of Contents CONT’D Number Title List of FiguresIii List of Tables List of Figures CONT’DSimple Menu Structure Section Description and SpecificationsHart Atex Recommended SensorsRecommended Ranges for Toroidal Sensors Transmitter Display During Calibration and Programming I T N E X THart Interface Devices Overview of Hart CommunicationFoundation Fieldbus AMS Main Menu Tools Orienting the Display Board Section InstallationMechanical Installation Unpacking and InspectionMounting on a Flat Surface Pipe Mounting 40308103Recommended Sensors Inductive LoopsSensor Selection Model 5081-T Sensor SelectionModel 5081-T-HT Power SUPPLY/CURRENT LoopPower Supply and Load Requirements. Refer to Figure Power Supply-Current Loop WiringTypical Fieldbus Network Electrical Wiring Configuration Power Supply Wiring for Model 5081-T-FF/FIWiring Through a Junction BOX Section WiringSensor Wiring Electrical Installation Sensor WiringPower Wiring Model 5081-T-FF Power Wiring Model 5081-T-HTWiring Model 242 sensor to Model 5081-T transmitter Electrical InstallationSection Section Section Intrinsically Safe & Explosion Proof IRC Infrared Remote ControlFM Explosion-Proof Installation for Model 5081-T-HT FM Intrinsically Safe Installation for Model 5081-T-HT CSA Intrinsically Safe Installation for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT Atex Intrisically Safe Label for Model 5081-T-HT FM Explosion-Proof Installation for Model 5081-T-FF FM Intrinsically Safe Installation for Model 5081-T-FF 10. CSA Intrinsically Safe Installation for Model 5081-T-FF 11. Atex Intrisically Safe Label for Model 5081-T-FF 12. Atex Intrinsically Safe Installation for Model 5081-T-FF FI-T 9241515-01 15. FM Intrinsically Safe Installation for Model 5081-T-FI 9241516-01 17. CSA Intrinsically Safe Installation for Model 5081-T-FI 18. Atex Intrisically Safe Label for Model 5081-T-FI 19. Atex Intrinsically Safe Installation for Model 5081-T-FI I T N E X T Section Display and OperationDisplays S UInfrared Remote Controller Infrared Remote Controller IRC KEY FunctionsQuick Start for Model 5081-T-HT Hart To reset transmiter to factory default settingsQuick Start for Model 5081-T-FF/FI Foundation Fieldbus Calibration Program Menu Tree Conductivity25.0C 12.00mA Process DisplayProgram Menu Mnemonics Default Settings Diagnostic MessagesDefault Settings for Model 5081-T-HT Retrieving a Lost Security Code Using HoldSecurity Entering the Security CodeSection START-UP and Calibration Accessing the Calibrate MenuModel 5081-T-FF Calibrate MenuOutput Cal Procedure9URU Temp SlopeU91 5ES-U1U17!91 GM!91General Section ProgramingModel 5081-T-HT Model 5081-T-FF Purpose Output 5081-HT onlyDefinitions Uhtu Procedure9VUVU 4QGTemp Process temperature is. Press Enter to saveU7-P H2SO4L DisplayG5-A HCLGgshtt Hart8VO!UT Setup CustPVO!RUT!!% SHIG12-U DefaultSetting Measurement Range Over Range Warning Limit RangePID Section Foundation Fieldbus OperationSection Operation with Model Connecting the CommunicatorOperation Making Hart related settings from the keypadOff-line and On-line Operation C/T-HT HART/Model 375 Menu Tree 1 PV LRV PV URVC/T-HT HART/Model 375 Menu Tree 2 LSL USLAO1 LRV URVC/T-HT HART/Model 375 Menu Tree 4 PV AO PV LRV PV URVMaxnotify Resource FaultstateBlockerr Featuresel Rsstate Faultstate LimnotifyTagdesc AlarmhysStrev BlockerrStrev Statusopts Strategy Alertkey HihilimPvftime LowcutIoopts LolimLolopri LololimBkcalhys Ffgain ResetRate Reset Sphilim SplolimBypass BaltimeAlertkey Outhilim ControloptsSphilim Strev Rate TagdescDvlopri DvhipriC/T-FF/FI Model 375 Menu Tree 9 C/T-FF/FI Model 375 Menu Tree 10 Overview Section Diagnosis and TroubleshootingCurrent Operating Menu Idvnut DLSTqiu 4DSGDisabling Faults Fault ConditionsNON- Disabling Warnings Hssqs 9WHS!S-PJ19IITHU!1SS Field Troubleshooting Quick Troubleshooting GuideQuick Troubleshooting Guide Symptom ActionTroubleshooting Flow Chart Systematic TroubleshootingRTD Resistance Values RTD Resistance ValuesFormula ExampleWhen a Fault exists When a Warning condition existsCPU Fail Troubleshooting When a Fault or Warning Message is ShowingTemp HI Temp LORtd FAIL, Temp HI, and Temp LO CPU Fail and ROM Fail 0 C 2 1 0 % mA Section MaintenancePreventative Maintenance Transmitter MaintenanceExploded View of Model 5081-T Transmitter Replacement Parts for Model 5081-T TransmitterOutput Logic Section Theory of OperationConductivity Hart CommunicationGeneral Warranty RepairNON-WARRANTY Repair Section Return of MaterialThis page intentionally left blank Return of Material WarrantySpecifications subject to change without notice