MODEL | SECTION 10.0 |
| DIAGNOSIS AND TROUBLESHOOTING |
10.5 SYSTEMATIC TROUBLESHOOTING
If the Quick Troubleshooting Guide does not resolve the error, try the
Step 1 Follow the troubleshooting flow chart.
Step 2 Refer to the tests and instructions indicated by the flow chart to diagnose the problem.
NOTE:
Before starting this procedure make sure that all wiring is correct.
NO
Conductivity Measurement
Problem (in the process)
Remove the sensor from process and place sensor in air. Zero instru- ment. Refer to Section 5.3 & 5.4.
OK?
YES
NOTE:
This step is for normal contact- ing only, not for low conductivity or resistivity.
Place sensor in process and
standardize. Refer to Section 5.3.
|
| YES |
|
| OK? |
|
| NO |
Check diagnostic |
|
|
messages | Remove sensor from process and test in known | |
| ||
Refer to Table | conductivity solution or against a certified | |
|
| conductivity instrument |
Check wiring | NO | YES |
for short |
| OK? |
Consult |
| YES |
Service Center |
|
|
Restart
Transmitter
Check for ground
loops and/or
improper installation
NO
Does problem
still exist?
FIGURE 10-4. Troubleshooting Flow Chart
80