5.0 Serial Communications

5.0 Serial Communications

The M460H comes equipped with a powerful digital Serial Communications Port that can be used for Data Acquisition and for changing instrument configuration. This port can be configured for either RS232 or RS485(half-duplex) operation. See Section 2.4.4 for details on connecting the port to a computer or Data Acquisition system.

5.1 Serial Port Configuration

The Serial Port of the device being used to communicate with the M460H should be configured as follows:

Table 5-1 Serial Port Configuration

Parameter

Value

Baud Rate

9600 bps

Data Bits

8

Stop Bits

1

Parity

None

5.2 Command Syntax

All characters sent and received are standard Ascii characters and all numbers are decimal numbers converted to Ascii text.

All commands are sent using the following syntax:

<address><command>:<data1>,<data2>#<checksum><CR>

Where:

address is the instrument address(default =1) command is the command string being sent

:(colon) is the data separator and is only included if data is being sent as part of the command (See Command Details below to see if a command requires data or not)

data1 is the first data parameter, if required.

data2 is the second data parameter, if required.

#is the Checksum separator, sent only if optional checksum is included checksum is an Ascii checksum of all characters up to the # character. The checksum is

optional, commands sent without the checksum(and checksum separator,) are also valid.

CR is a carriage return, Ascii 13.

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 21

Image 21
Contents INSTRUCTION MANUAL MODEL 460H NEMA OZONE MONITOREMAIL api-sales@teledyne.com 03662 Rev. D 1/8/04SAFETY MESSAGES 2.0 INSTALLATION TABLE OF CONTENTS6.0 MAINTENANCE I. 0 INTRODUCTION7.0 SENSOR AND SYSTEM TROUBLESHOOTING 6.2 CLEANING EXTERIOR SURFACES OF THE M460HAPPENDIX A - SERIAL PORT CONFIGURATION 6.3 DEGREE OF PROTECTIONFigure 2-1 Mounting Hole Locations FIGURESPneumatic Connections Electrical ConnectionsTable 2-1 Digital Status Output Pins TABLESTable 2-3 Serial Port Pins Serial Port Configuration1.1 Preface I.0 INTRODUCTIONADVANCED POLLUTION INSTRUMENTATION DIVISION 1.2 WARRANTY POLICY02024c TERMS AND CONDITIONS1.3 Principle of Operation 1.4 Specifications 2.0 Installation 2.1 Mechanical InstallationFigure 2-1 Mounting Hole Locations Figure 2-2 Pneumatic Connections 2.2 Pneumatic Connections2.3AC Power Connection 2.4.1 Analog Output 2.4 Electrical I/O ConnectionsFigure 2-3 Electrical Connections I/O CONNECTOR2.4.2 Digital Status Outputs Table 2-1 Digital Status Output PinsLabel NameTable 2-2 Control Inputs 2.4.3 Control InputsInput Description2.4.4 RS232/RS485 Serial Communications Port Table 2-3 Serial Port PinsFigure 2-5 Typical RS232 and RS485 Connections RS2323.0 Startup 4.1 Front Panel Display 4.0 Operation4.1.1 O3 readout 4.1.2 Pressure Display4.2.3 Auto Zero Error 4.2.1 Auto-Zero Installation4.2 Auto-Zero Operation 4.2.2 Auto-Zero Function5.2 Command Syntax 5.1 Serial Port ConfigurationTable 5-1 Serial Port Configuration addresscommanddata1,data2#checksumCRExamples of valid Commands with no data 5.3 Command SummaryExamples of valid Commands with data Table 5-2 Serial Port Command Summary5.4.1 CAUTO 5.4 Command ReferenceSyntax Data Parameters Sent5.4.2 CZERO 5.4.3 DACSTEP 5.4.4 O3 addressO3CRaddresso3conc#checksumCR addressSETADDRnewaddressCR 5.4.5 SETADDR5.4.6 TDUMP 5.4.7 TLIST Allowable Range 5.4.8 VGETIndex addressvarvalue#checksumCR5.4.9 VLIST 1VLISTCR#5 iirfilt = 0.4 #6 concunits = varindex 5.4.10 VSETis index number for internal VAR See VGET for index list newvalue6.2 Cleaning Exterior Surfaces of the M460H 6.0 MaintenanceFilter Replacement Procedure 6.1 Replacing the gas filter element7.1 Status Output Summary 7.0 Sensor and System Troubleshooting7.2 Sensor OK Table 7-1 Digital Status Outputs7.3 Invalid Reading 7.4 Lamp Low7.5 Cell Dirty Table 7-2 Status Output Truth Table 7.6 Status Output Summary TableSensor InvalidAppendix A - Serial Port Configuration Figure A-1 RS232/RS485 Jumpers Controller PCA#03492RS232RS485 Appendix B - Pneumatic Diagram