7.6 Status Output Summary Table

7.6 Status Output Summary Table

Table 7-2 below is a logic truth table summarizing the recommended actions based on the states of the four status outputs. A ‘1’ indicates the output is ON, a ‘0’ indicates the output is OFF, and ‘X’ indicates the output is in either state.

Table 7-2: Status Output Truth Table

Sensor

Invalid

Lamp

Cell

Actions

OK

Reading

Low

Dirty

 

1

0

0

0

Normal operation, no action

 

 

 

 

required

0

X

X

X

Service required

1

1

X

X

Check Pressure > 45 psia

 

 

 

 

Verify that concentration has

 

 

 

 

not exceeded full scale range

 

 

 

 

of sensor.

 

 

 

 

Calibrate at Zero.

1

X

1

X

Lamp adjustment useful,

 

 

 

 

though not required

1

X

X

1

Calibrate at zero

 

 

 

 

Clean Cell

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 36

Image 36
Contents MODEL 460H NEMA OZONE MONITOR INSTRUCTION MANUALEMAIL api-sales@teledyne.com 03662 Rev. D 1/8/04SAFETY MESSAGES TABLE OF CONTENTS 2.0 INSTALLATION6.0 MAINTENANCE I. 0 INTRODUCTION6.2 CLEANING EXTERIOR SURFACES OF THE M460H 7.0 SENSOR AND SYSTEM TROUBLESHOOTINGAPPENDIX A - SERIAL PORT CONFIGURATION 6.3 DEGREE OF PROTECTIONFIGURES Figure 2-1 Mounting Hole LocationsPneumatic Connections Electrical ConnectionsTABLES Table 2-1 Digital Status Output PinsTable 2-3 Serial Port Pins Serial Port ConfigurationI.0 INTRODUCTION 1.1 Preface1.2 WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION02024c TERMS AND CONDITIONS1.3 Principle of Operation 1.4 Specifications 2.0 Installation 2.1 Mechanical InstallationFigure 2-1 Mounting Hole Locations 2.2 Pneumatic Connections Figure 2-2 Pneumatic Connections2.3AC Power Connection 2.4 Electrical I/O Connections 2.4.1 Analog OutputFigure 2-3 Electrical Connections I/O CONNECTORTable 2-1 Digital Status Output Pins 2.4.2 Digital Status OutputsLabel Name2.4.3 Control Inputs Table 2-2 Control InputsInput DescriptionTable 2-3 Serial Port Pins 2.4.4 RS232/RS485 Serial Communications PortFigure 2-5 Typical RS232 and RS485 Connections RS2323.0 Startup 4.0 Operation 4.1 Front Panel Display4.1.1 O3 readout 4.1.2 Pressure Display4.2.1 Auto-Zero Installation 4.2.3 Auto Zero Error4.2 Auto-Zero Operation 4.2.2 Auto-Zero Function5.1 Serial Port Configuration 5.2 Command SyntaxTable 5-1 Serial Port Configuration addresscommanddata1,data2#checksumCR5.3 Command Summary Examples of valid Commands with no dataExamples of valid Commands with data Table 5-2 Serial Port Command Summary5.4 Command Reference 5.4.1 CAUTOSyntax Data Parameters Sent5.4.2 CZERO 5.4.3 DACSTEP 5.4.4 O3 addressO3CRaddresso3conc#checksumCR 5.4.5 SETADDR addressSETADDRnewaddressCR5.4.6 TDUMP 5.4.7 TLIST 5.4.8 VGET Allowable RangeIndex addressvarvalue#checksumCR5.4.9 VLIST 1VLISTCR#5 iirfilt = 0.4 #6 concunits = 5.4.10 VSET varindexis index number for internal VAR See VGET for index list newvalue6.0 Maintenance 6.2 Cleaning Exterior Surfaces of the M460HFilter Replacement Procedure 6.1 Replacing the gas filter element7.0 Sensor and System Troubleshooting 7.1 Status Output Summary7.2 Sensor OK Table 7-1 Digital Status Outputs7.3 Invalid Reading 7.4 Lamp Low7.5 Cell Dirty 7.6 Status Output Summary Table Table 7-2 Status Output Truth TableSensor InvalidAppendix A - Serial Port Configuration Figure A-1 RS232/RS485 Jumpers Controller PCA#03492RS232RS485 Appendix B - Pneumatic Diagram