5.4.8 VGET

Syntax

<address>VGET:<var_index><CR>

Description

Returns value of an internal configuration variable (VAR.)

Data Parameters Sent

var_index is index number for internal VAR as follows:

Var

Name

Description

 

Allowable Range

Index

 

 

 

 

0

analog_range

Full-Scale concentration

5.0 – 400.0 (Units are

 

 

range for Analog Output

current unit of measure,

 

 

scaling

wt% or g/Nm3)

1

azero_enable

Auto-Zero Enable VAR.

0

= Off

 

 

(only valid if Auto-Zero

1

= On

 

 

hardware is installed)

 

 

2

azero_period

Period for Auto-Zero

5.0 - 86400.0 Seconds

 

 

Calibration timer

 

 

3

carrier_weight

Mol weight of carrier gas

27.0 – 32.0 mol weight

 

 

for wt% calculations (32.0

 

 

 

 

= O2)

 

 

4

comm_mode

Not Used

N/A

5

iir_filt

Digital Concentration

0.05 – 1.0

 

 

filter

(1.0 = No Filtering, lower

 

 

 

values increases filtering)

6

conc_units

O3 Concentration

0

= wt%

 

 

measurement units

1

= g/Nm3

Response

<address>:<var_value>#<checksum><CR>

Example

Command:

1VGET:0<CR> - Request analog_range value

Response:

1:15.0#303 - analog_range returned as 15.0 (wt%)

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 30

Image 30
Contents EMAIL api-sales@teledyne.com MODEL 460H NEMA OZONE MONITORINSTRUCTION MANUAL 03662 Rev. D 1/8/04SAFETY MESSAGES 6.0 MAINTENANCE TABLE OF CONTENTS2.0 INSTALLATION I. 0 INTRODUCTIONAPPENDIX A - SERIAL PORT CONFIGURATION 6.2 CLEANING EXTERIOR SURFACES OF THE M460H7.0 SENSOR AND SYSTEM TROUBLESHOOTING 6.3 DEGREE OF PROTECTIONPneumatic Connections FIGURESFigure 2-1 Mounting Hole Locations Electrical ConnectionsTable 2-3 Serial Port Pins TABLESTable 2-1 Digital Status Output Pins Serial Port ConfigurationI.0 INTRODUCTION 1.1 Preface02024c 1.2 WARRANTY POLICYADVANCED POLLUTION INSTRUMENTATION DIVISION TERMS AND CONDITIONS1.3 Principle of Operation 1.4 Specifications 2.0 Installation 2.1 Mechanical InstallationFigure 2-1 Mounting Hole Locations 2.2 Pneumatic Connections Figure 2-2 Pneumatic Connections2.3AC Power Connection Figure 2-3 Electrical Connections 2.4 Electrical I/O Connections2.4.1 Analog Output I/O CONNECTORLabel Table 2-1 Digital Status Output Pins2.4.2 Digital Status Outputs NameInput 2.4.3 Control InputsTable 2-2 Control Inputs DescriptionFigure 2-5 Typical RS232 and RS485 Connections Table 2-3 Serial Port Pins2.4.4 RS232/RS485 Serial Communications Port RS2323.0 Startup 4.1.1 O3 readout 4.0 Operation4.1 Front Panel Display 4.1.2 Pressure Display4.2 Auto-Zero Operation 4.2.1 Auto-Zero Installation4.2.3 Auto Zero Error 4.2.2 Auto-Zero FunctionTable 5-1 Serial Port Configuration 5.1 Serial Port Configuration5.2 Command Syntax addresscommanddata1,data2#checksumCRExamples of valid Commands with data 5.3 Command SummaryExamples of valid Commands with no data Table 5-2 Serial Port Command SummarySyntax 5.4 Command Reference5.4.1 CAUTO Data Parameters Sent5.4.2 CZERO 5.4.3 DACSTEP 5.4.4 O3 addressO3CRaddresso3conc#checksumCR 5.4.5 SETADDR addressSETADDRnewaddressCR5.4.6 TDUMP 5.4.7 TLIST Index 5.4.8 VGETAllowable Range addressvarvalue#checksumCR5.4.9 VLIST 1VLISTCR#5 iirfilt = 0.4 #6 concunits = is index number for internal VAR See VGET for index list 5.4.10 VSETvarindex newvalueFilter Replacement Procedure 6.0 Maintenance6.2 Cleaning Exterior Surfaces of the M460H 6.1 Replacing the gas filter element7.2 Sensor OK 7.0 Sensor and System Troubleshooting7.1 Status Output Summary Table 7-1 Digital Status Outputs7.3 Invalid Reading 7.4 Lamp Low7.5 Cell Dirty Sensor 7.6 Status Output Summary TableTable 7-2 Status Output Truth Table InvalidAppendix A - Serial Port Configuration Figure A-1 RS232/RS485 Jumpers Controller PCA#03492RS232RS485 Appendix B - Pneumatic Diagram