Teledyne 460H Setaddr, addressSETADDRnewaddressCR, Syntax, Description, Data Parameters Sent

Page 27

5.4.5 SETADDR

Syntax

<address>SETADDR:<new_address><CR>

Description

Changes the communications address to a new value.

Data Parameters Sent

new_address is the new address for the monitor; Range for new_address is 1-9.

Response

<address>:<success_flag>#<checksum><CR>

Example

Command:

1SETADDR:2<CR> - Change address from 1 to 2

Response:

1:OK#261<CR> - Change Address Successful 1:FAIL#391<CR> - Change Address Failed

Notes

Note that monitor response is from previous address, but any further commands must be at new address. In the example above, after sending back the OK response, the monitor will no longer respond to commands with address 1.

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 27

Image 27
Contents 03662 Rev. D 1/8/04 MODEL 460H NEMA OZONE MONITORINSTRUCTION MANUAL EMAIL api-sales@teledyne.comSAFETY MESSAGES I. 0 INTRODUCTION TABLE OF CONTENTS2.0 INSTALLATION 6.0 MAINTENANCE6.3 DEGREE OF PROTECTION 6.2 CLEANING EXTERIOR SURFACES OF THE M460H7.0 SENSOR AND SYSTEM TROUBLESHOOTING APPENDIX A - SERIAL PORT CONFIGURATIONElectrical Connections FIGURESFigure 2-1 Mounting Hole Locations Pneumatic ConnectionsSerial Port Configuration TABLESTable 2-1 Digital Status Output Pins Table 2-3 Serial Port Pins1.1 Preface I.0 INTRODUCTIONTERMS AND CONDITIONS 1.2 WARRANTY POLICYADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c1.3 Principle of Operation 1.4 Specifications 2.0 Installation 2.1 Mechanical InstallationFigure 2-1 Mounting Hole Locations Figure 2-2 Pneumatic Connections 2.2 Pneumatic Connections2.3AC Power Connection I/O CONNECTOR 2.4 Electrical I/O Connections2.4.1 Analog Output Figure 2-3 Electrical ConnectionsName Table 2-1 Digital Status Output Pins2.4.2 Digital Status Outputs LabelDescription 2.4.3 Control InputsTable 2-2 Control Inputs InputRS232 Table 2-3 Serial Port Pins2.4.4 RS232/RS485 Serial Communications Port Figure 2-5 Typical RS232 and RS485 Connections3.0 Startup 4.1.2 Pressure Display 4.0 Operation4.1 Front Panel Display 4.1.1 O3 readout4.2.2 Auto-Zero Function 4.2.1 Auto-Zero Installation4.2.3 Auto Zero Error 4.2 Auto-Zero Operationaddresscommanddata1,data2#checksumCR 5.1 Serial Port Configuration5.2 Command Syntax Table 5-1 Serial Port ConfigurationTable 5-2 Serial Port Command Summary 5.3 Command SummaryExamples of valid Commands with no data Examples of valid Commands with dataData Parameters Sent 5.4 Command Reference5.4.1 CAUTO Syntax5.4.2 CZERO 5.4.3 DACSTEP 5.4.4 O3 addressO3CRaddresso3conc#checksumCR addressSETADDRnewaddressCR 5.4.5 SETADDR5.4.6 TDUMP 5.4.7 TLIST addressvarvalue#checksumCR 5.4.8 VGETAllowable Range Index5.4.9 VLIST 1VLISTCR#5 iirfilt = 0.4 #6 concunits = newvalue 5.4.10 VSETvarindex is index number for internal VAR See VGET for index list6.1 Replacing the gas filter element 6.0 Maintenance6.2 Cleaning Exterior Surfaces of the M460H Filter Replacement ProcedureTable 7-1 Digital Status Outputs 7.0 Sensor and System Troubleshooting7.1 Status Output Summary 7.2 Sensor OK7.3 Invalid Reading 7.4 Lamp Low7.5 Cell Dirty Invalid 7.6 Status Output Summary TableTable 7-2 Status Output Truth Table SensorAppendix A - Serial Port Configuration Figure A-1 RS232/RS485 Jumpers Controller PCA#03492RS232RS485 Appendix B - Pneumatic Diagram