Teledyne 460H instruction manual Invalid Reading, Lamp Low, Cell Dirty

Page 35
7.3 Invalid Reading

7.3 Invalid Reading

The normal state for the Invalid Reading output is OFF. If this output turns on, this indicates that the M460H is still operational, but a system fault or calibration fault exists that may make the current ozone reading invalid.

The Invalid Reading output is turned on for any of the following conditions:

1.When the measured pressure in the measurement cell exceeds 45 psia.

2.When the measured concentration has exceeded the full-scale concentration range of the sensor. Check the serial number tag for the full-scale concentration range.

3.The sensor is indicating an excessive negative reading.

7.4Lamp Low

The normal state for the Lamp Low output is OFF. If this output turns on, this indicates that the UV lamp intensity as measured by the reference detector has dropped below 2500mV.

If the Lamp Low output turns ON and the Sensor O.K. output is ON, this indicates that the lamp intensity is still adequate for measurement, but adjustment should be made when possible.

If the Lamp Low output turns ON and the Sensor O.K. output is OFF, this indicates a failure condition and measurement is no longer possible.

7.5 Cell Dirty

The normal state for the Cell Dirty output is OFF. If this output turns on, then the ratio of the measure detector to the reference detector (at zero) is < 0.5. This value is calculated when the zero calibration is performed.

When this output is on, it indicates a loss of optical transmission through the windows in the absorption cell or a calibration fault.

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 35

Image 35
Contents 03662 Rev. D 1/8/04 MODEL 460H NEMA OZONE MONITORINSTRUCTION MANUAL EMAIL api-sales@teledyne.comSAFETY MESSAGES I. 0 INTRODUCTION TABLE OF CONTENTS2.0 INSTALLATION 6.0 MAINTENANCE6.3 DEGREE OF PROTECTION 6.2 CLEANING EXTERIOR SURFACES OF THE M460H7.0 SENSOR AND SYSTEM TROUBLESHOOTING APPENDIX A - SERIAL PORT CONFIGURATIONElectrical Connections FIGURESFigure 2-1 Mounting Hole Locations Pneumatic ConnectionsSerial Port Configuration TABLESTable 2-1 Digital Status Output Pins Table 2-3 Serial Port Pins1.1 Preface I.0 INTRODUCTIONTERMS AND CONDITIONS 1.2 WARRANTY POLICYADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c1.3 Principle of Operation 1.4 Specifications Figure 2-1 Mounting Hole Locations 2.0 Installation2.1 Mechanical Installation Figure 2-2 Pneumatic Connections 2.2 Pneumatic Connections2.3AC Power Connection I/O CONNECTOR 2.4 Electrical I/O Connections2.4.1 Analog Output Figure 2-3 Electrical ConnectionsName Table 2-1 Digital Status Output Pins2.4.2 Digital Status Outputs LabelDescription 2.4.3 Control InputsTable 2-2 Control Inputs InputRS232 Table 2-3 Serial Port Pins2.4.4 RS232/RS485 Serial Communications Port Figure 2-5 Typical RS232 and RS485 Connections3.0 Startup 4.1.2 Pressure Display 4.0 Operation4.1 Front Panel Display 4.1.1 O3 readout4.2.2 Auto-Zero Function 4.2.1 Auto-Zero Installation4.2.3 Auto Zero Error 4.2 Auto-Zero Operationaddresscommanddata1,data2#checksumCR 5.1 Serial Port Configuration5.2 Command Syntax Table 5-1 Serial Port ConfigurationTable 5-2 Serial Port Command Summary 5.3 Command SummaryExamples of valid Commands with no data Examples of valid Commands with dataData Parameters Sent 5.4 Command Reference5.4.1 CAUTO Syntax5.4.2 CZERO 5.4.3 DACSTEP addresso3conc#checksumCR 5.4.4 O3addressO3CR addressSETADDRnewaddressCR 5.4.5 SETADDR5.4.6 TDUMP 5.4.7 TLIST addressvarvalue#checksumCR 5.4.8 VGETAllowable Range Index#5 iirfilt = 0.4 #6 concunits = 5.4.9 VLIST1VLISTCR newvalue 5.4.10 VSET varindex is index number for internal VAR See VGET for index list6.1 Replacing the gas filter element 6.0 Maintenance6.2 Cleaning Exterior Surfaces of the M460H Filter Replacement ProcedureTable 7-1 Digital Status Outputs 7.0 Sensor and System Troubleshooting7.1 Status Output Summary 7.2 Sensor OK7.5 Cell Dirty 7.3 Invalid Reading7.4 Lamp Low Invalid 7.6 Status Output Summary TableTable 7-2 Status Output Truth Table SensorRS232RS485 Appendix A - Serial Port ConfigurationFigure A-1 RS232/RS485 Jumpers Controller PCA#03492 Appendix B - Pneumatic Diagram