13. If the O2 cell current is too strong or too weak to span, an error message will appear. This will not occur if the user-entered span gas concentration is correct. Press ACK to retry, or ESC to abort the span.

14. If the span is successful, the system will return the main Analyze mode menu. Check to see that the span value is correct.

4-10

TELEDYNE ANALYTICAL INSTRUMENTS

Image 38
Contents Model Copyright 1999 Teledyne Analytical Instruments Iii Table of Contents Maintenance & Troubleshooting AppendixTeledyne Analytical Instruments Variable Features FeaturesFixed Features IntroductionIntroduction 3160DA Dual fuel cell Introduction ComponentsOptions and Model Numbers Introduction Applications Operational Theory Principles of OperationMajor Components Operational Theory Micro-Fuel Cell Sensor Sample SystemOperational Theory Electronics and Signal Processing Model 3160 flow schematicOperational Theory Analog signal output offset Identifier Range Scale Voltage CurrentmADefaults Programmable Options Front Panel InstallationUnpacking the Analyzer Qty DescriptionRear Panel Analyzer front and rear panels Instrument Air Compressed Air Fitting Installation Gas Line ConnectionsSpan Gas Sample GasSensor Installation Electrical Connections Voltage Selection 3160 Rear PanelFuse Changing Removing the Voltage CardFuse Replacement Installation Checklist 3 RS-232 Serial Digital PortOperations Front Panel ControlsCell Output Factor Span Factor Modes of OperationSpinning Wheel Operations Changing Passwords For Remote Monitoring and Control Uled Span in ProgressOperations Cold Start-Up Operations Calibration Zeroing Operations Operations Calibration Using Span Gas Operations Operations Operations Select Active Sensor Two Cells Only Setup Operations Operations O2 Range Set-Up AccptOperations Set-Up Alarms Cold StartOperations Operations Logger Set-Up Operations Operations Set-Up Clock Functions Accpt Teledyne Analytical Instruments Operations Operations System Statistics Comm. Std. for 1200B Bell USA/CCITT Baud rateAscii Data Bits 7 or Stop Bits 1 or Operations Operations Sensor Maintenance Maintenance & TroubleshootingRoutine Maintenance Scrubber MaintenanceTroubleshooting Symptom Cause CorrectionMaintenance & Troubleshooting Symptom Cause Correction Maintenance & Troubleshooting Appendix SpecificationsOperating Temperature0-50C 32 -122 F Altitude 1,609 m Relative Humidy Up to 99% Teledyne Analytical InstrumentsAppendix Recommended Spare Parts List Drawing ListDate Prepared 2/12/96 Appendix Material Safety Data SheetCustomer Service Extension Environmental Health Material or Component S. # QuantitySection IV Fire and Explosion Hazard Data Protective measures during cell replacement ExposureSection Viii Control Measures Address Product NameManufacturer PhoneExposure limits Osha PEL Signs/symptoms of exposurePrimary route of entry Effects of overexposure IngestionProtective measures Disposal