4 Operations
Calibration Using Span Gas
1. Enter the Calibration mode from the Analyze mode by pressing SPAN.
2. Press ACK if zeroing has already been done. Make sure to zero before calibrating to ensure correct O2 readings.
3. If the span gas tank is not connected to the span port, connect it now and press ACK.
4. Enter the span gas concentration using the UP and DOWN keys, and press ACCPT.
5. Set the flow rate to 2 SCFH and press ACK.
6. Select manual or automatic sensor settling detection by pressing MAN or AUTO. Automatic sensor settling de- tection lets the system determine when the sensor has stabilized.