Operations 4
System Statistics
1.Enter
SETUP.
ESC in any status page will take you to the last
PREV steps you back one status page. NEXT moves forward one status page.
2.Press STAT for system status.
(If only one cell block is installed, TEST and INSTL will be omitted.)
3.Status page one displays the active sensor (A or B), the system time/date, and the system hardware configura- tion. The number and date of the installed software ver- sion are also shown.
Press PREV for the previous screen, or NEXT for the next screen.
4.Status page two shows the O2 level set for spanning, the cell strength factor, the type and time of the last span, the time interval for scheduled spanning (AutoSpan), and whether or not scheduled spanning is active. [The cell strength factor is an indication of the amount of cell life left in the cell, with 1.00 being the most and 0.00 be- ing the least.]
5.Status page three shows the time of the last zero, the time interval for scheduled zeroing (AutoZero), and whether or not scheduled zeroing is active.
6.Status page four shows the settings for the five pro- grammable alarms.
TELEDYNE ANALYTICAL INSTRUMENTS |