Intelligent Motion Systems UDAS-1001E user manual Analog Output Test

Page 30

Checking the System

3.3.2.2 Analog Output Test

The analog output test controls the voltage on the analog output channels of the UDAS unit.

In the UDAS SYSCHECK Test Category Selection Window, click the Analog Out button and the UDAS SYSCHECK ANALOG OUTPUT dialog shown in FIGURE 3.4, UDAS SYSCHECK Analog Output Test of a UDAS unit opens.

In the Channel 0 and Channel 1 fields, click to insert the cursor and enter the desired value (volts or counts) in any of the four data entry fields.

After replacing the old value with a new value, press the Enter key to update both analog outputs from the UDAS unit.

Clicking the Disable Outputs check box grounds both of the analog output channels.

FIGURE 3.4 UDAS SYSCHECK Analog Output Test of a UDAS unit

3-6

Image 30
Contents UDAS-1001E Series Warranty and Repair Policy Statement GeneralOnline Support Email Support Technical Assistance and ServiceStatic Sensitivity North American Customers International CustomersTrademarks Use of EquipmentFCC Radio Frequency Interference Statement Revision History Table of Contents Specifications Introduction Features SoftwareSupport LibraryAnalog Input Analog OutputDigital Input/Output Udas 1001E Series Functional DescriptionData Acquisition Functions CounterHigh-Speed Analog Input Mode Starting and Stopping the High-Speed Analog Input ModeStarting Conversions Rate Generator Channel ScannerStopping Conversions High-Speed Analog Input Triggering Auto Zero Digital I/OExample Counters UDAS-1001E-4 dimensional drawing Diagnostic Software Support and AccessoriesPower Software SupportVisual Designer Udas Support Library LabVIEW Udas Support LibraryPower-on Default Digital I/O Configuration Removing the CoverSetting the Configuration Udas Switch S1Reinstalling the Cover Enabling the CJCDifferential JP1 Jumper Position JP2 Jumper Position Connecting the Udas Unit Making the Initial ConnectionChecking the System Use your CD drive letter\Syscheck\SETUP.EXEENTER Installing and Running Udas SyscheckInstalling Syscheck Syscheck Operation Unit Info System Check Analog Input Test Analog Output Test Counter Test Udas Syscheck Counter Test Measure Input High Mode Udas Syscheck Counter Test Measure Frequency Mode Digital I/O Test Udas Syscheck Digital I/O TestPaced Analog Input 10 High-Speed Aquisition Sample Troubleshooting Tips USB PortPower Test Troubleshooting ErrorsUdas Drivers Check the wiring and cables for shorts or damage Specifications UDAS-1001E Series Hardware SpecificationsRMS USB UDAS-1001E-3, UDAS-1001E-4 Dimensions Figure A.1 Dimensions of the Udas unit Top viewFigure A.3 Dimensions of the Udas unit Bottom view Figure A.4 Dimensional Drawing of the Udas series PCB Udas I/O Connectors Udas I/O Connectors Udas I/O Connectors Accessories Udas Termination Panels PCI-20429T-1 Multifunction Termination PanelFigure B.2 PCI-20429T-1 Multifunction Termination Panel Udas Termination and Signal Conditioning Options A L O G T E R M I N a T I O N SFigure B.4 UDAS-1001E Termination Options Analog In Single-Ended and Differential Function Summary of Features Isolator BlocksPCI-1100 Series Digital Opto-Isolator Blocks Udas Analog Input Circuit Calibration Procedures Removing the Udas Unit’s CoverEquipment Required Analog Input and Output Calibration Procedures Analog Input Calibration Procedure Adjust Offset and Gain for ±10Volt Input RangeAdjust Input Offset Voltage Adjust Offset and Gain for ± 5Volt Input Range Adjust Offset and Gain for 0 to 10 Volt Range Udas Analog Output Circuit Calibration Procedure Analog Output Channel 0 Offset and Gain Analog Output Channel 1 Offset and GainIndex Udas Index-3