Agilent Technologies 66332A, 6634B, 6633B Triggering Output Changes, Scpi Triggering Nomenclature

Page 21

Programming the DC Source - 3

Triggering Output Changes

The dc source has two independent trigger systems. One is used for generating output changes, and the other is used for triggering measurements. This section describes the output trigger system. The measurement trigger system is described under "Triggering Measurements".

SCPI Triggering Nomenclature

In SCPI terms, trigger systems are called sequences. When more than one trigger system exists, they are differentiated by naming them SEQuence1 and SEQuence2. SEQuence1 is the transient trigger system and SEQuence2 is the measurement trigger system. The dc source uses aliases with more descriptive names for these sequences. These aliases can be used instead of the sequence forms.

Sequence Form

Alias

SEQuence1

TRANsient

SEQuence2

ACQuire

Output Trigger System Model

Figure 3-1 is a model of the output trigger system. The rectangular boxes represent states. The arrows show the transitions between states. These are labeled with the input or event that causes the transition to occur.

INITiate:CONTinuous OFF

INITiate:CONTinuous ON

 

ABORt

IDLE STATE

*RST

 

*RCL

INITiate[:IMMediate]

INITIATED STATE

TRIGGER RECEIVED

OUTPUT

LEVEL

CHANGE

Figure 3-1. Model of Output Triggers

Setting the Voltage or Current Trigger Levels

To program output trigger levels, you must first specify a voltage or current trigger level that the output will go to once a trigger signal is received. Use the following commands to set the output trigger level:

VOLTage:TRIGgered <n> or

CURRent:TRIGgered <n>

NOTE: Until they are programmed, uninitialized trigger levels will assume their corresponding immediate levels. For example, if a dc source is powered up and VOLTage:LEVel is programmed to 6, then VOLTage:LEVel:TRIGger will also be 6 until you program it to another value. Once you program VOLTage:LEVel:TRIGger to a value, it will remain at that value regardless of how you subsequently reprogram VOLTage:LEVel.

21

Image 21
Contents Programming Guide Printing History Safety GuidelinesTable of Contents Internally Triggered Measurements Language DictionaryOutput Commands Scpi Conformance Information Error MessagesCompatibility Language Example ProgramsDocumentation Summary About this GuideExternal References Gpib ReferencesScpi References Downloading and Installing the Driver VXIplug&play Power Products Instrument DriversSupported Applications System RequirementsGpib Capabilities of the DC Source Accessing Online HelpRS-232 Capabilities of the DC Source Gpib AddressRS-232 Flow Control RS-232 Programming ExampleBaud Rate Introduction to Scpi RS-232 TroubleshootingConventions Used in This Guide Boldface fontMultiple Commands in a Message Types of Scpi CommandsUsing Queries Including Common CommandsTypes of Scpi Messages Moving Among SubsystemsMessage Unit Query IndicatorHeaders Message Unit SeparatorNumerical Data Formats Scpi Data FormatsSuffixes and Multipliers Response Data TypesUsing Device Clear Scpi Command CompletionPage Introduction Power-on InitializationProgramming the Output Enabling the OutputOutput Current Output VoltageMaximum Voltage Maximum CurrentTriggering Output Changes Setting the Voltage or Current Trigger LevelsScpi Triggering Nomenclature Output Trigger System ModelGenerating Triggers Initiating the Output Trigger SystemMaking Measurements Voltage and Current MeasurementsDC Measurements Current Ranges RMS Measurements Agilent 66312A, 66332A OnlyInternally Triggered Measurements Measurement Trigger System ModelSequence Form Alias SEQuence2 ACQuire BUS Generating Measurement Triggers Agilent 66312A, 66332A OnlyTrigger Commands Used to Measure Output Pulses Measuring Output Pulses Agilent 66312A, 66332A Only Current DetectorPulse Measurement Queries Controlling Measurement Samples Varying the Voltage or Current Sampling RateMultiple Measurements Agilent 66312A, 66332A Only Pre-event and Post-event Triggering Pulse Measurement Example Agilent 66312A, 66332A onlyOption Base Programming the Status Registers Power-On ConditionsOperation Status Group Questionable Status Group Standard Event Status GroupStatus Byte Register Servicing Operation Status and Questionable Status Events Determining the Cause of a Service InterruptMSS Bit RQS BitMonitoring Both Phases of a Status Transition Inhibit/Fault IndicatorDiscrete Fault Indicator DFI Remote Inhibit RIBit Weight Using the Inhibit/Fault Port as a Digital I/ODFI Programming Example PinPage Language Dictionary Subsystem CommandsSubsystem Commands Syntax ALC DFISubsystem Commands Syntax Common Commands Syntax Common CommandsProgramming Parameters Output Programming ParametersCALibrateCURRent Calibration CommandsCALibrateCURRentNEGative CALibrateCURRentMEASureLOWRangeCommand Syntax CALibrateSAVE Parameters None Examples CALibratePASSwordCALibrateDATA CALibrateLEVelCALibrateVOLTage CALibrateSTATeCALibrateVOLTagePROTection Query Syntax CALibrateSTATe?MEASureARRayCURRent? FETChARRayCURRent? Measurement CommandsMEASureARRayVOLTage? FETChARRayVOLTage? Query SyntaxMEASureCURRentACDC? FETChCURRentACDC? MEASureCURRent? FETChCURRent?MEASureCURRentHIGH? FETChCURRentHIGH? FETChCURRent? applies to Agilent 66312A, 66332A OnlyMEASureCURRentLOW? FETChCURRentLOW? MEASureCURRentMAXimum? FETChCURRent MAXimum?MEASureCURRentMINimum? FETChCURRentMINimum? MEASureVOLTageACDC? FETChVOLTageACDC? MEASureVOLTage? FETChVOLTage?MEASureVOLTageHIGH? FETChVOLTageHIGH? FETChVOLTage? applies to Agilent 66312A, 66332A OnlyMEASureVOLTageLOW? FETChVOLTageLOW? MEASureVOLTageMAXimum? FETChVOLTageMAXimum?MEASureVOLTageMINimum? FETChVOLTageMINimum? SENSeCURRentDETector SENSeCURRentRANGeUnit a amperes *RST Value MAX high range Returned Parameters NR3SENSeSWEepOFFSetPOINts SENSeFUNCtionSENSeSWEepPOINts SENSeSWEepTINTervalHANNing SENSeWINDowRECTangular Returned Parameters CRDOUTPut Output CommandsOUTPutDFI OUTPutDFISOURceOUTPutPROTectionDELay OUTPutPONSTATeOUTPutPROTectionCLEar OUTPutRELayPOLarity OUTPutRELayOUTPutRIMODE Query Syntax OUTPputRELayPOLarity?SOURceCURRent Default SuffixSOURceCURRentTRIGger SOURceCURRentPROTectionSTATeSOURceDIGitalFUNCtion SOURceDIGitalDATASOURceVOLTage Query Syntax SOURceDIGitalDATA?SOURceVOLTageTRIGger SOURceVOLTageALCBANDwidth? SOURceVOLTageALCBWIDth?SOURceVOLTagePROTection Agilent 66332A, 6631B, 6632B, 6633B and 6634B OnlySTATusPRESet Status CommandsSTATusOPERation? STATusOPERationCONDition?Query Syntax STATusOPERationENABle? STATusOPERationENABleParameters Preset Value STATusOPERationNTR STATusOPERationPTRSTATusQUEStionableENABle STATusQUEStionable?STATusQUEStionableCONDition? STATusQUEStionableNTR STATusQUEStionablePTR Command Syntax *CLS Parameters NoneCLS STATQUESNTR?STATQUESPTR?ESE Bit Configuration of Standard Event Status Enable RegisterESR? OPCSRE PSCPSC on ExampleSTB? Bit Configuration of Status Byte RegisterWAI NR1 register binary valueDISPlay System CommandsDISPlayMODE DISPlayTEXTSYSTemLANGuage SYSTemERRor?SYSTemVERSion? Parameters none Returned Parameters NR2SYSTemREMote SYSTemLOCalSYSTemRWLock IDN?OPT? Command Syntax RCL NRf Parameters ExampleRCL RSTCommand Syntax SAV NRf Parameters Example RST SettingsSAV TST?ABORt Trigger CommandsINITiateSEQuence INITiateNAME INITiateCONTinuousSEQuence1 INITiateCONTinuousNAMETRIGgerSOURce TRIGgerTRIGgerSEQuence2 TRIGgerACQuire Abor Currtrig Init *TRG VolttrigTRIGgerSEQuence2COUNtVOLTage TRIGgerACQuireCOUNtVOLTage TRIGgerSEQuence2COUNtCURRent TRIGgerACQuireCOUNtCURRentParameters RST Value Examples TRIGSEQ2COUNCURR 5 TrigacqcouncurrTRIGSEQ2HYSTCURR TRIGSEQ2LEVVOLT TRIGSEQ2HYSTVOLT TRIGSEQ2LEVCURRTRIGgerSEQuence2LEVelVOLTage TRIGgerACQuireLEVelVOLTage TRIGgerSEQuence2LEVelCURRent TRIGgerACQuireLEVelCURRentTRIGSEQ2LEVVOLT TRIGSEQ2HYSTCURR TRIGSEQ2LEVCURR TRIGSEQ2HYSTVOLTTRIGgerSEQuence2SLOPeCURRent TRIGgerACQuireSLOPeCURRent TRIGgerSEQuence2SLOPeVOLTage TRIGgerACQuireSLOPeVOLTageTRIGSEQ2SLOPVOLT TRIGgerSEQuence2SOURce TRIGgerACQuireSOURce Parameters None Related CommandsTRIGgerSEQuence1DEFine TRIGgerSEQuence2DEFine TRGPage Scpi Confirmed Commands Non-SCPI CommandsScpi Version Page Command Setting Table B-1. COMPatibility Power-on Settings CommandERR? FAULT? ID? IOUT? Table B-2. COMPatibility Commands CompatibilitySimilar Scpi Command ASTS? CLRTable B-2. COMPatibility Commands Description Similar Scpi Compatibility CommandCompatibility Description Command Table B-2. COMPatibility CommandsThese commands determine the conditions that will set bits ErrorNorm Fast INH ERR UNR +CC RQS ERR RDYFAU Page Bit Set Error Number Error Code Error Type Error Number ListTable C-1. Error Numbers Error Number Page Types of DOS Drivers Assigning the Gpib Address in ProgramsAgilent 82335A Driver National Instruments Gpib DriverBasic Controllers Error HandlingExample Programs D Call IBCLRPS% Example 2. IBM Controller Using National InterfaceCODES$=*CLS Option Base Example 3. Controller Using BasicIndex 100 IndexScpi 101102 Manual Updates
Related manuals
Manual 82 pages 25.5 Kb Manual 83 pages 60.43 Kb

6613C, 66312A, 6631B, 6611C, 6614C specifications

Agilent Technologies, a leader in electronic test and measurement solutions, offers a range of power supplies designed to meet various application needs. Notable models include the 6632B, 6634B, 66332A, 6633B, and 6612C. Each of these units provides unique features and technologies that cater to researchers, engineers, and technicians in the industry.

The Agilent 6632B is a single-output DC power supply that delivers up to 30V and 3A. It is known for its excellent load regulation and low noise, making it ideal for sensitive electronic testing. The model includes built-in voltage and current measurement capabilities, allowing users to monitor output conditions in real time. The 6632B is commonly used in laboratory environments, educational institutions, and manufacturing lines.

Moving to the 6634B, this model offers dual-output capabilities with a maximum output of 30V and 6A. This versatility enables simultaneous powering of two different devices or circuit sections. It also features parallel and series operation options, allowing users to create a custom power supply configuration for specific applications. With a programmable interface, the 6634B simplifies test automation, ensuring efficiency in extensive testing scenarios.

The Agilent 66332A stands out with its precision and high performance. This power supply provides three outputs—two programmable and one fixed—yielding flexible power configurations. Its intuitive user interface allows easy adjustment of voltage and current settings. The device is equipped with extensive protection features to safeguard both the power supply and the connected load against faults. It is an excellent choice for complex testing setups that require reliable power.

The 6633B model offers a high-performance power supply with dual outputs, similar to the 6634B but with enhanced specifications. It can provide up to 40V and 2A per channel, delivering precision for demanding applications. This model is particularly suited for industries focused on high-reliability applications, such as telecommunications and aerospace.

Lastly, the Agilent 6612C is a compact and lightweight power supply providing single-output up to 60V and 2A. This model is designed for simplicity and ease of use, making it an excellent choice for portable applications. The 6612C’s unique characteristics include a compact design and user-friendly controls, which facilitate operation in field settings.

In summary, Agilent Technologies’ power supply models—6632B, 6634B, 66332A, 6633B, and 6612C—offer an array of features that cater to a wide range of testing and research needs, ensuring reliable power delivery in various contexts.