Agilent Technologies 6632B, 6634B, 66332A, 6633B, 6614C, 6613C, 6631B, 66312A Operation Status Group

Page 33

Bit

0

5

8

10

11

0

1

2

4

9

10

14

0

2

3

4

5

7

3

4

5

6

7

Programming the DC Source - 3

Table 3-1. Bit Configurations of Status Registers

Signal

Meaning

CAL

Operation Status Group

The dc source is computing new calibration constants

WTG

The dc source is waiting for a trigger

CV

The dc source is in constant voltage mode

CC+

The dc source is in constant current mode

CC-

The dc source is in negative constant current mode

OV

Questionable Status Group

The overvoltage protection has tripped

OCP

The overcurrent protection has tripped

FS

The fuse is blown

OT

The overtemperature protection has tripped

RI

The remote inhibit state is active

Unreg

The output is unregulated

MeasOvld

Current measurement exceeded capability of low range

OPC

Standard Event Status Group

Operation complete

QYE

Query error

DDE

Device-dependent error

EXE

Execution error

CME

Command error

PON

Power-on

QUES

Status Byte and Service Request Enable Registers

Questionable status summary bit

MAV

Message Available summary bit

ESB

Event Status Summary bit

MSS

Master Status Summary bit

RQS

Request Service bit

OPER

Operation status summary bit

Operation Status Group

The Operation Status registers record signals that occur during normal operation. As shown below, the group consists of a Condition, PTR/NTR, Event, and Enable register. The outputs of the Operation Status register group are logically-ORed into the OPER(ation) summary bit (7) of the Status Byte register.

Register

Condition

PTR Filter

NTR Filter

Event

Enable

Command

STAT:OPER:COND?

STAT:OPER:PTR <n>

STAT:OPER:NTR <n>

STAT:OPER:EVEN?

STAT:OPER:ENAB <n>

Description

A register that holds real-time status of the circuits being monitored. It is a read-only register.

A positive transistion filter that functions as described under STAT:OPER:NTRPTR commands in chapter 4. It is a read/write register.

A negative transition filter that functions as described under STAT:OPER:NTRPTR commands in chapter 4. It is a read/write register.

A register that latches any condition that is passed through the PTR or NTR filters. It is a read-only register that is cleared when read.

A register that functions as a mask for enabling specific bits from the Event register. It is a read/write register.

33

Image 33
Contents Programming Guide Printing History Safety GuidelinesTable of Contents Internally Triggered Measurements Language DictionaryOutput Commands Scpi Conformance Information Error MessagesCompatibility Language Example ProgramsDocumentation Summary About this GuideExternal References Gpib ReferencesScpi References Downloading and Installing the Driver VXIplug&play Power Products Instrument DriversSupported Applications System RequirementsGpib Capabilities of the DC Source Accessing Online HelpRS-232 Capabilities of the DC Source Gpib AddressRS-232 Flow Control RS-232 Programming ExampleBaud Rate Introduction to Scpi RS-232 TroubleshootingConventions Used in This Guide Boldface fontMultiple Commands in a Message Types of Scpi CommandsUsing Queries Including Common CommandsTypes of Scpi Messages Moving Among SubsystemsMessage Unit Query IndicatorHeaders Message Unit SeparatorNumerical Data Formats Scpi Data FormatsSuffixes and Multipliers Response Data TypesUsing Device Clear Scpi Command CompletionPage Introduction Power-on InitializationProgramming the Output Enabling the OutputOutput Current Output VoltageMaximum Voltage Maximum CurrentTriggering Output Changes Setting the Voltage or Current Trigger LevelsScpi Triggering Nomenclature Output Trigger System ModelGenerating Triggers Initiating the Output Trigger SystemMaking Measurements Voltage and Current MeasurementsDC Measurements Current Ranges RMS Measurements Agilent 66312A, 66332A OnlyInternally Triggered Measurements Measurement Trigger System ModelSequence Form Alias SEQuence2 ACQuire BUS Generating Measurement Triggers Agilent 66312A, 66332A OnlyTrigger Commands Used to Measure Output Pulses Measuring Output Pulses Agilent 66312A, 66332A Only Current DetectorPulse Measurement Queries Controlling Measurement Samples Varying the Voltage or Current Sampling RateMultiple Measurements Agilent 66312A, 66332A Only Pre-event and Post-event Triggering Pulse Measurement Example Agilent 66312A, 66332A onlyOption Base Programming the Status Registers Power-On ConditionsOperation Status Group Questionable Status Group Standard Event Status GroupStatus Byte Register Servicing Operation Status and Questionable Status Events Determining the Cause of a Service InterruptMSS Bit RQS BitMonitoring Both Phases of a Status Transition Inhibit/Fault IndicatorDiscrete Fault Indicator DFI Remote Inhibit RIBit Weight Using the Inhibit/Fault Port as a Digital I/ODFI Programming Example PinPage Language Dictionary Subsystem CommandsSubsystem Commands Syntax ALC DFISubsystem Commands Syntax Common Commands Syntax Common CommandsProgramming Parameters Output Programming ParametersCALibrateCURRent Calibration CommandsCALibrateCURRentNEGative CALibrateCURRentMEASureLOWRangeCommand Syntax CALibrateSAVE Parameters None Examples CALibratePASSwordCALibrateDATA CALibrateLEVelCALibrateVOLTage CALibrateSTATeCALibrateVOLTagePROTection Query Syntax CALibrateSTATe?MEASureARRayCURRent? FETChARRayCURRent? Measurement CommandsMEASureARRayVOLTage? FETChARRayVOLTage? Query SyntaxMEASureCURRentACDC? FETChCURRentACDC? MEASureCURRent? FETChCURRent?MEASureCURRentHIGH? FETChCURRentHIGH? FETChCURRent? applies to Agilent 66312A, 66332A OnlyMEASureCURRentLOW? FETChCURRentLOW? MEASureCURRentMAXimum? FETChCURRent MAXimum?MEASureCURRentMINimum? FETChCURRentMINimum? MEASureVOLTageACDC? FETChVOLTageACDC? MEASureVOLTage? FETChVOLTage?MEASureVOLTageHIGH? FETChVOLTageHIGH? FETChVOLTage? applies to Agilent 66312A, 66332A OnlyMEASureVOLTageLOW? FETChVOLTageLOW? MEASureVOLTageMAXimum? FETChVOLTageMAXimum?MEASureVOLTageMINimum? FETChVOLTageMINimum? SENSeCURRentDETector SENSeCURRentRANGeUnit a amperes *RST Value MAX high range Returned Parameters NR3SENSeSWEepOFFSetPOINts SENSeFUNCtionSENSeSWEepPOINts SENSeSWEepTINTervalHANNing SENSeWINDowRECTangular Returned Parameters CRDOUTPut Output CommandsOUTPutDFI OUTPutDFISOURceOUTPutPROTectionDELay OUTPutPONSTATeOUTPutPROTectionCLEar OUTPutRELayPOLarity OUTPutRELayOUTPutRIMODE Query Syntax OUTPputRELayPOLarity?SOURceCURRent Default SuffixSOURceCURRentTRIGger SOURceCURRentPROTectionSTATeSOURceDIGitalFUNCtion SOURceDIGitalDATASOURceVOLTage Query Syntax SOURceDIGitalDATA?SOURceVOLTageTRIGger SOURceVOLTageALCBANDwidth? SOURceVOLTageALCBWIDth?SOURceVOLTagePROTection Agilent 66332A, 6631B, 6632B, 6633B and 6634B OnlySTATusPRESet Status CommandsSTATusOPERation? STATusOPERationCONDition?Query Syntax STATusOPERationENABle? STATusOPERationENABleParameters Preset Value STATusOPERationNTR STATusOPERationPTRSTATusQUEStionableENABle STATusQUEStionable?STATusQUEStionableCONDition? STATusQUEStionableNTR STATusQUEStionablePTR Command Syntax *CLS Parameters NoneCLS STATQUESNTR?STATQUESPTR?ESE Bit Configuration of Standard Event Status Enable RegisterESR? OPCSRE PSCPSC on ExampleSTB? Bit Configuration of Status Byte RegisterWAI NR1 register binary valueDISPlay System CommandsDISPlayMODE DISPlayTEXTSYSTemLANGuage SYSTemERRor?SYSTemVERSion? Parameters none Returned Parameters NR2SYSTemREMote SYSTemLOCalSYSTemRWLock IDN?OPT? Command Syntax RCL NRf Parameters ExampleRCL RSTCommand Syntax SAV NRf Parameters Example RST SettingsSAV TST?ABORt Trigger CommandsINITiateSEQuence INITiateNAME INITiateCONTinuousSEQuence1 INITiateCONTinuousNAMETRIGgerSOURce TRIGgerTRIGgerSEQuence2 TRIGgerACQuire Abor Currtrig Init *TRG VolttrigTRIGgerSEQuence2COUNtVOLTage TRIGgerACQuireCOUNtVOLTage TRIGgerSEQuence2COUNtCURRent TRIGgerACQuireCOUNtCURRentParameters RST Value Examples TRIGSEQ2COUNCURR 5 TrigacqcouncurrTRIGSEQ2HYSTCURR TRIGSEQ2LEVVOLT TRIGSEQ2HYSTVOLT TRIGSEQ2LEVCURRTRIGgerSEQuence2LEVelVOLTage TRIGgerACQuireLEVelVOLTage TRIGgerSEQuence2LEVelCURRent TRIGgerACQuireLEVelCURRentTRIGSEQ2LEVVOLT TRIGSEQ2HYSTCURR TRIGSEQ2LEVCURR TRIGSEQ2HYSTVOLTTRIGgerSEQuence2SLOPeCURRent TRIGgerACQuireSLOPeCURRent TRIGgerSEQuence2SLOPeVOLTage TRIGgerACQuireSLOPeVOLTageTRIGSEQ2SLOPVOLT TRIGgerSEQuence2SOURce TRIGgerACQuireSOURce Parameters None Related CommandsTRIGgerSEQuence1DEFine TRIGgerSEQuence2DEFine TRGPage Scpi Confirmed Commands Non-SCPI CommandsScpi Version Page Command Setting Table B-1. COMPatibility Power-on Settings CommandERR? FAULT? ID? IOUT? Table B-2. COMPatibility Commands CompatibilitySimilar Scpi Command ASTS? CLRTable B-2. COMPatibility Commands Description Similar Scpi Compatibility CommandCompatibility Description Command Table B-2. COMPatibility CommandsThese commands determine the conditions that will set bits ErrorNorm Fast INH ERR UNR +CC RQS ERR RDYFAU Page Bit Set Error Number Error Code Error Type Error Number ListTable C-1. Error Numbers Error Number Page Types of DOS Drivers Assigning the Gpib Address in ProgramsAgilent 82335A Driver National Instruments Gpib DriverBasic Controllers Error HandlingExample Programs D Call IBCLRPS% Example 2. IBM Controller Using National InterfaceCODES$=*CLS Option Base Example 3. Controller Using BasicIndex 100 IndexScpi 101102 Manual Updates
Related manuals
Manual 82 pages 25.5 Kb Manual 83 pages 60.43 Kb

6613C, 66312A, 6631B, 6611C, 6614C specifications

Agilent Technologies, a leader in electronic test and measurement solutions, offers a range of power supplies designed to meet various application needs. Notable models include the 6632B, 6634B, 66332A, 6633B, and 6612C. Each of these units provides unique features and technologies that cater to researchers, engineers, and technicians in the industry.

The Agilent 6632B is a single-output DC power supply that delivers up to 30V and 3A. It is known for its excellent load regulation and low noise, making it ideal for sensitive electronic testing. The model includes built-in voltage and current measurement capabilities, allowing users to monitor output conditions in real time. The 6632B is commonly used in laboratory environments, educational institutions, and manufacturing lines.

Moving to the 6634B, this model offers dual-output capabilities with a maximum output of 30V and 6A. This versatility enables simultaneous powering of two different devices or circuit sections. It also features parallel and series operation options, allowing users to create a custom power supply configuration for specific applications. With a programmable interface, the 6634B simplifies test automation, ensuring efficiency in extensive testing scenarios.

The Agilent 66332A stands out with its precision and high performance. This power supply provides three outputs—two programmable and one fixed—yielding flexible power configurations. Its intuitive user interface allows easy adjustment of voltage and current settings. The device is equipped with extensive protection features to safeguard both the power supply and the connected load against faults. It is an excellent choice for complex testing setups that require reliable power.

The 6633B model offers a high-performance power supply with dual outputs, similar to the 6634B but with enhanced specifications. It can provide up to 40V and 2A per channel, delivering precision for demanding applications. This model is particularly suited for industries focused on high-reliability applications, such as telecommunications and aerospace.

Lastly, the Agilent 6612C is a compact and lightweight power supply providing single-output up to 60V and 2A. This model is designed for simplicity and ease of use, making it an excellent choice for portable applications. The 6612C’s unique characteristics include a compact design and user-friendly controls, which facilitate operation in field settings.

In summary, Agilent Technologies’ power supply models—6632B, 6634B, 66332A, 6633B, and 6612C—offer an array of features that cater to a wide range of testing and research needs, ensuring reliable power delivery in various contexts.