Anritsu 0701012, MA24106A manual Remote Operation Command Details

Page 38

Remote Operation Command Details

Remote Operation

4-3 Remote Operation Command Details

Each command needs to be followed by a Line Feed (0x0A, ASCII 10) termination character. Response from the sensor will have a Line Feed (0x0A, ASCII 10) termination character attached at the end.

START

Description: Puts the power sensor in measurement mode. Syntax: START + LF

Return Value: None for the first time, OK for any subsequent command sent.

Remarks: This command does not return anything when sent the first time. For any subsequent START commands, the sensor will return OK. This is helpful, if the user wants to know if the sensor is in measurement mode.

STOP

Description: Stops the measurement mode and puts the sensor in idle mode.

Syntax: STOP + LF

Return Value: OK or ERR

Remarks: This command should be sent before exiting the user application.

IDN?

Description: Gets identification information from the sensor.

Syntax: IDN? + LF

Return Value: ANRITSU, Model #, Serial #, Module Serial #, firmware version

PWR?

Description: Gets the power reading from the power sensor. Syntax: PWR? + LF

Return Value: Power value in dBm

Remarks: If an error condition exists, the returned power reading values are pre-tagged with the letter “E” and the sensor’s LED turns yellow. The STATUS? command can then be issued to find details about the error.

NPWR?

Description: Gets a new power reading from the power sensor. Syntax: NPWR? + LF

Return Value: Current Power value in dBm

Remarks: After receiving this command, the power sensor discards the existing data that is stored in the sensor’s buffer. A new measurement of the current power is initiated to get a new power reading.

If an error condition exists, the returned power reading values are pre-tagged with the letter “E” and the sensor’s LED turns yellow. The STATUS? command can then be issued to find details about the error.

Note: This command is only available in sensor firmware versions 1.01 and later.

4-2

MA24106A UG

Image 38
Contents MA24106A USB Power Sensor User Guide True-RMS, 50 MHz to 6 GHzUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentHardware and Software Requirements InstallationDriver Installation Driver Installation Anritsu Power Meter InstallationInstallation Driver Installation License AgreementSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Using the Power Sensor Graphical User InterfaceData Entry Fields Buttons Using the Power SensorButtons Status Bar Using the Power Sensor Display WindowDisplay Window File Menu Menu Bar Using the Power SensorMenu Bar Tools MenuData Logging Menu DataLogging MenuPower Graph 10. Data LoggingTime, X-axis 30 minutes Offset Table 14.Offset Entry Screen16. Open Dialog MA24106A UG Session Restore 17.Offset Table Menu Offset Table Off StatusUsing the Power Sensor Making Measurements Basic Power MeasurementConnecting the DUT Making MeasurementsCalibrating the Sensor Making Measurements Using the Power SensorZeroing the Sensor Applying a Calibration Factor CorrectionNumber Averages Needed for ±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm NoiseMeasurement Considerations Using the Power Sensor Error StatesTime Varying Signals Message Description ResolutionUsing the Power Sensor Measurement Considerations High Crest Factor Signals peak to average ratioMultitone Signals Noise and AveragingSettling Time Using the Power Sensor Uncertainty of a Measurement Uncertainty of a MeasurementUncertainty Example Uncertainty of a Measurement Using the Power SensorUncertainty Adjusted Specification Noise Calculations at 12 dBm 16 mWRemote Operation Remote Operation Commands SummaryCommand Description Remote Operation Command Details Remote Operation Command DetailsRemote Operation Remote Operation Command Details FREQ?HAT Sensor Operational Tests Required EquipmentVswr Pretest Sensor Operational Tests Test ProcedureVswr Pretest Calibration Factor Test Sensor Operational Tests Calibration Factor TestCalibration Factor Test Sensor Operational Tests MA24106ALinearity Test Sensor Operational Tests Linearity TestLinearity Test Sensor Operational Tests ApplySensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Appendix a Connector Care and Handling Beware of destructive pin depth mating connectorsCleaning connectors Avoid over torquing connectorsAvoid applying excessive power Connector Care and HandlingDemo Application Using the Demo ApplicationAppendix B Sample Visual Basic Code Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine For Anritsu Power Meter 1.0 Only Appendix C Serial Port CompatibilityMethod 1-Download Updated Software Method 3-Remapping a Serial PortSerial Port Compatibility Figure C-2.System PropertiesMethod 3-Remapping a Serial Port Figure C-3.Device ManagerFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Appendix D Upgrading the Firmware Upgrading the FirmwareFigure D-3.Firmware Upgrade Dialog Upgrading the FirmwareFigure D-5.Open File Dialog Figure D-7 Index SymbolsSymbols to L To S Index-2 MA24106A UGTo Z USB