Calibration Factor Test | Sensor Operational Tests |
2.Connect the reference sensor to the synthesizer with the appropriate adapter and attenuator
3.Apply the appropriate Cal factor to the reference sensor per the manufacturer’s instruction.
4.Record the power indicated by the reference meter in the appropriate space in Table
5.Disconnect the reference sensor from the synthesizer output and connect the MA24106A power sensor with the appropriate adapter and attenuator
6.Apply the appropriate Cal factor to the MA24106A as follows:
Press the Frequency button on the Power Meter application, and then enter the frequency of the measurement in GHz.
7.Record the power indicated by the MA24106A in the appropriate space in Table
8.Set the synthesizer frequency to the next frequency in Table
9.Repeat steps 2 through 8 until all of the frequencies in Table
10.For each row in Table
11.For each frequency, compare the power difference to the maximum allowed difference specified in Table
Table
| A | B |
| |
| Reference Power | MA24106A | Absolute Value of |
|
| Difference in Power | Maximum Allowed | ||
Frequency | Measurement | Measurement | Measurements | Difference |
(GHz) | (dBm) | (dBm) | (dB) | (dB) |
|
|
|
|
|
0.05 |
|
|
| 0.26 |
|
|
|
|
|
0.1 |
|
|
| 0.26 |
|
|
|
|
|
0.3 |
|
|
| 0.26 |
|
|
|
|
|
0.5 |
|
|
| 0.26 |
|
|
|
|
|
1.0 |
|
|
| 0.26 |
|
|
|
|
|
2.0 |
|
|
| 0.31 |
|
|
|
|
|
3.0 |
|
|
| 0.31 |
|
|
|
|
|
4.0 |
|
|
| 0.31 |
|
|
|
|
|
5.0 |
|
|
| 0.33 |
|
|
|
|
|
6.0 |
|
|
| 0.33 |
|
|
|
|
|
MA24106A UG |