Anritsu MA24106A, 0701012 manual Sensor Operational Tests Linearity Test

Page 47

Sensor Operational Tests

Linearity Test

7.Set up the test for the second 20 dB range as follows:

a.Remove the 10 dB attenuator from in between the reference sensor and splitter and connect the reference sensor directly to the splitter.

b.Remove the MA24106A from the splitter and connect the 10 dB attenuator between the splitter and the MA24106A power sensor (see Figure 5-3).

c.Turn Off the synthesizer’s RF output and perform a low-level Zero of both the Reference sensor and the MA24106A.

d.Turn On the synthesizer’s RF output.

Figure 5-3.Linearity Test Setup

e.Set the synthesizer output level to +10 dBm and then adjust it until the sensor/meter under test reads as close as possible to the value obtained above in step 6e.

8.Record data for the next 20 dB range

a.Read and record the power indicated by the reference meter in the appropriate place in Table 5-4.

b.Lower the output power level of the synthesizer to +5 dBm.

c.Record the reference meter and the MA24106A power sensor readings in the appropriate place in Table 5-4.

MA24106A UG

5-7

Image 47
Contents True-RMS, 50 MHz to 6 GHz MA24106A USB Power Sensor User GuideUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentHardware and Software Requirements InstallationDriver Installation Anritsu Power Meter Installation Driver InstallationLicense Agreement Installation Driver InstallationSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Graphical User Interface Using the Power SensorData Entry Fields Buttons Using the Power SensorButtons Status Bar Using the Power Sensor Display WindowDisplay Window Tools Menu Menu Bar Using the Power SensorMenu Bar File MenuDataLogging Menu Data Logging Menu10. Data Logging Power GraphTime, X-axis 30 minutes 14.Offset Entry Screen Offset Table16. Open Dialog MA24106A UG 17.Offset Table Menu Offset Table Off Status Session RestoreMaking Measurements Basic Power MeasurementConnecting the DUT Using the Power Sensor Making MeasurementsApplying a Calibration Factor Correction Making Measurements Using the Power SensorZeroing the Sensor Calibrating the Sensor±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm Noise Number Averages Needed forMessage Description Resolution Error StatesTime Varying Signals Measurement Considerations Using the Power SensorNoise and Averaging High Crest Factor Signals peak to average ratioMultitone Signals Using the Power Sensor Measurement ConsiderationsSettling Time Uncertainty of a Measurement Using the Power Sensor Uncertainty of a MeasurementNoise Calculations at 12 dBm 16 mW Uncertainty of a Measurement Using the Power SensorUncertainty Adjusted Specification Uncertainty ExampleRemote Operation Remote Operation Commands SummaryCommand Description Remote Operation Command Details Remote Operation Command DetailsFREQ? Remote Operation Remote Operation Command DetailsHAT Required Equipment Sensor Operational TestsVswr Pretest Sensor Operational Tests Test ProcedureVswr Pretest Sensor Operational Tests Calibration Factor Test Calibration Factor TestMA24106A Calibration Factor Test Sensor Operational TestsSensor Operational Tests Linearity Test Linearity TestApply Linearity Test Sensor Operational TestsSensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Beware of destructive pin depth mating connectors Appendix a Connector Care and HandlingConnector Care and Handling Avoid over torquing connectorsAvoid applying excessive power Cleaning connectorsDemo Application Using the Demo ApplicationAppendix B Sample Visual Basic Code Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine Appendix C Serial Port Compatibility For Anritsu Power Meter 1.0 OnlyMethod 3-Remapping a Serial Port Method 1-Download Updated SoftwareFigure C-2.System Properties Serial Port CompatibilityFigure C-3.Device Manager Method 3-Remapping a Serial PortFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Upgrading the Firmware Appendix D Upgrading the FirmwareUpgrading the Firmware Figure D-3.Firmware Upgrade DialogFigure D-5.Open File Dialog Figure D-7 Index SymbolsSymbols to L Index-2 MA24106A UG To SUSB To Z