Anritsu 0701012, MA24106A manual Hat

Page 40

Remote Operation Command Details

Remote Operation

HAT

Description: Sets the high aperture time mode.

Syntax: HAT + LF

Return Value: OK or ERR

Remarks: This command will put the sensor in high aperture time mode. In this mode, the A to D converter integration time is about 160 milliseconds.

LAT

Description: Sets the low aperture time mode.

Syntax: HAT + LF

Return Value: OK or ERR

Remarks: This command will put the sensor in low aperture time mode. In this mode, the A to D converter integration time is about 10 milliseconds. This mode is the default mode for the sensor when powered up.

4-4

MA24106A UG

Image 40
Contents MA24106A USB Power Sensor User Guide True-RMS, 50 MHz to 6 GHzUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentDriver Installation InstallationHardware and Software Requirements Driver Installation Anritsu Power Meter InstallationInstallation Driver Installation License AgreementSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Using the Power Sensor Graphical User InterfaceButtons Buttons Using the Power SensorData Entry Fields Display Window Using the Power Sensor Display WindowStatus Bar Menu Bar Using the Power Sensor Menu BarFile Menu Tools MenuData Logging Menu DataLogging MenuPower Graph 10. Data LoggingTime, X-axis 30 minutes Offset Table 14.Offset Entry Screen16. Open Dialog MA24106A UG Session Restore 17.Offset Table Menu Offset Table Off StatusBasic Power Measurement Connecting the DUTUsing the Power Sensor Making Measurements Making MeasurementsMaking Measurements Using the Power Sensor Zeroing the SensorCalibrating the Sensor Applying a Calibration Factor CorrectionNumber Averages Needed for ±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm NoiseError States Time Varying SignalsMeasurement Considerations Using the Power Sensor Message Description ResolutionHigh Crest Factor Signals peak to average ratio Multitone SignalsUsing the Power Sensor Measurement Considerations Noise and AveragingSettling Time Using the Power Sensor Uncertainty of a Measurement Uncertainty of a MeasurementUncertainty of a Measurement Using the Power Sensor Uncertainty Adjusted SpecificationUncertainty Example Noise Calculations at 12 dBm 16 mWCommand Description Remote Operation Commands SummaryRemote Operation Remote Operation Command Details Remote Operation Command DetailsRemote Operation Remote Operation Command Details FREQ?HAT Sensor Operational Tests Required EquipmentVswr Pretest Test ProcedureVswr Pretest Sensor Operational Tests Calibration Factor Test Sensor Operational Tests Calibration Factor TestCalibration Factor Test Sensor Operational Tests MA24106ALinearity Test Sensor Operational Tests Linearity TestLinearity Test Sensor Operational Tests ApplySensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Appendix a Connector Care and Handling Beware of destructive pin depth mating connectorsAvoid over torquing connectors Avoid applying excessive powerCleaning connectors Connector Care and HandlingAppendix B Sample Visual Basic Code Using the Demo ApplicationDemo Application Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine For Anritsu Power Meter 1.0 Only Appendix C Serial Port CompatibilityMethod 1-Download Updated Software Method 3-Remapping a Serial PortSerial Port Compatibility Figure C-2.System PropertiesMethod 3-Remapping a Serial Port Figure C-3.Device ManagerFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Appendix D Upgrading the Firmware Upgrading the FirmwareFigure D-3.Firmware Upgrade Dialog Upgrading the FirmwareFigure D-5.Open File Dialog Figure D-7 Symbols to L SymbolsIndex To S Index-2 MA24106A UGTo Z USB