Anritsu 0701012, MA24106A manual Test Procedure, Vswr Pretest Sensor Operational Tests

Page 42

VSWR Pretest

Sensor Operational Tests

5-3 VSWR Pretest

The most common cause of power sensor failure is excess input power. Applying power exceeding the damage level shown on the label will damage the sensor’s sensing element resulting in impedance change. Input match will be degraded when element impedance is changed. If you suspect that a senor is damaged, you should start with an input match pretest.

The maximum VSWR values are listed in the Performance Specification section of this manual. The uncertainty of the VSWR test equipment will affect actual measurement values. See the following Table 5-2 for an example of how measurement system uncertainty can affect the Expected Maximum Reflection Coefficient when using the Anritsu MS4624B Vector Network Analyzer.

Test Procedure

Follow the manufacturers S11 (or return loss) calibration procedure to perform calibration on a network analyzer. Connect the power sensor to the network analyzer test port and measure power sensor input match. Typically, matches are measured in terms of return loss in dB. Return loss and magnitude of the reflection coefficient conversion equations are as follows:

ρ= 10–RL/20

RL = –20logρwhere

RL = Return Loss in dB

ρ= Magnitude of the Reflection Coefficient

VSWR and magnitude of the reflection coefficient conversion equations are as follows: VSWR = (1 + ρ) / (1 – ρ)

ρ= (VSWR – 1) / (VSWR + 1) where

VSWR = Voltage Standing Wave Ratio

ρ= Magnitude of the Reflection Coefficient

Record the measured data into Table 5-2 under the Actual Measurement column. The Actual Measurement should be smaller than the Maximum Reflection coefficient. The Maximum Reflection Coefficient is equal to the measurement system uncertainty added to the sensor’s reflection coefficient specification. If the Actual Measurement reflection coefficient is larger than the Maximum Reflection Coefficient, then the power sensor may be defective. If the actual reflection coefficient is significantly larger than the maximum values in Table 5-2, then the sensor is damaged and it is not necessary to perform further testing.

Note: There are no user-serviceable parts inside the power sensors. Contact your local Anritsu Service Center and return defective sensors with a detailed description of the observed problem.

Table 5-2.Pretest Measurement Result

 

MS4624B Reflection

Maximum Reflection

 

Frequency

Coefficient Uncertainty

Coefficient

Actual Measurement

 

 

 

 

50 MHz to 2 GHz

0.013

0.050 + 0.013 = 0.063

 

 

 

 

 

2 GHz to 6 GHz

0.020

0.100 + 0.020 = 0.120

 

 

 

 

 

5-2

MA24106A UG

Image 42
Contents MA24106A USB Power Sensor User Guide True-RMS, 50 MHz to 6 GHzUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentInstallation Driver InstallationHardware and Software Requirements Driver Installation Anritsu Power Meter InstallationInstallation Driver Installation License AgreementSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Using the Power Sensor Graphical User InterfaceButtons Using the Power Sensor ButtonsData Entry Fields Using the Power Sensor Display Window Display WindowStatus Bar File Menu Menu Bar Using the Power SensorMenu Bar Tools MenuData Logging Menu DataLogging MenuPower Graph 10. Data LoggingTime, X-axis 30 minutes Offset Table 14.Offset Entry Screen16. Open Dialog MA24106A UG Session Restore 17.Offset Table Menu Offset Table Off StatusUsing the Power Sensor Making Measurements Basic Power MeasurementConnecting the DUT Making MeasurementsCalibrating the Sensor Making Measurements Using the Power SensorZeroing the Sensor Applying a Calibration Factor CorrectionNumber Averages Needed for ±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm NoiseMeasurement Considerations Using the Power Sensor Error StatesTime Varying Signals Message Description ResolutionUsing the Power Sensor Measurement Considerations High Crest Factor Signals peak to average ratioMultitone Signals Noise and AveragingSettling Time Using the Power Sensor Uncertainty of a Measurement Uncertainty of a MeasurementUncertainty Example Uncertainty of a Measurement Using the Power SensorUncertainty Adjusted Specification Noise Calculations at 12 dBm 16 mWRemote Operation Commands Summary Command DescriptionRemote Operation Remote Operation Command Details Remote Operation Command DetailsRemote Operation Remote Operation Command Details FREQ?HAT Sensor Operational Tests Required EquipmentTest Procedure Vswr PretestVswr Pretest Sensor Operational Tests Calibration Factor Test Sensor Operational Tests Calibration Factor TestCalibration Factor Test Sensor Operational Tests MA24106ALinearity Test Sensor Operational Tests Linearity TestLinearity Test Sensor Operational Tests ApplySensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Appendix a Connector Care and Handling Beware of destructive pin depth mating connectorsCleaning connectors Avoid over torquing connectorsAvoid applying excessive power Connector Care and HandlingUsing the Demo Application Appendix B Sample Visual Basic CodeDemo Application Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine For Anritsu Power Meter 1.0 Only Appendix C Serial Port CompatibilityMethod 1-Download Updated Software Method 3-Remapping a Serial PortSerial Port Compatibility Figure C-2.System PropertiesMethod 3-Remapping a Serial Port Figure C-3.Device ManagerFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Appendix D Upgrading the Firmware Upgrading the FirmwareFigure D-3.Firmware Upgrade Dialog Upgrading the FirmwareFigure D-5.Open File Dialog Figure D-7 Symbols Symbols to LIndex To S Index-2 MA24106A UGTo Z USB