Method 1–Download Updated Software

Serial Port Compatibility

C-1 Method 1–Download Updated Software

The preferred method for resolving serial port compatibility issues is to download software updates for your product from www.us.anritsu.com.

C-2 Method 2–Trying a Different USB Port

1.Disconnect the USB end of your power sensor from your computer (or USB hub).

2.Connect the USB power sensor to a different USB port on your computer. Connecting to a USB hub tends to increase the port numbers, so connecting directly to the computer’s USB port usually provides the best result.

3.Open the Device manager to see if the new port assignment is between COM1 and COM16. If it is NOT, return to step 1 and connect to a different USB port. If the port assignment is between COM1 and COM16, the problem has been resolved and no further action is required.

C-3 Method 3–Remapping a Serial Port

A serial port may be remapped to a different number, such as changing a serial port from COM19 to COM5. This may be needed if method 1 above does not result in a serial port assignment between COM1 and COM16 or if you prefer to use a USB hub or a specific USB port on your computer.

1.Open the System Properties by going to Start Settings Control Panel System, or simultaneously pressing the Windows and Pause keys.

C-2

MA24106A UG

Image 58
Contents MA24106A USB Power Sensor User Guide True-RMS, 50 MHz to 6 GHzUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentDriver Installation InstallationHardware and Software Requirements Driver Installation Anritsu Power Meter InstallationInstallation Driver Installation License AgreementSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Using the Power Sensor Graphical User InterfaceButtons Buttons Using the Power SensorData Entry Fields Display Window Using the Power Sensor Display WindowStatus Bar File Menu Menu Bar Using the Power SensorMenu Bar Tools MenuData Logging Menu DataLogging MenuPower Graph 10. Data LoggingTime, X-axis 30 minutes Offset Table 14.Offset Entry Screen16. Open Dialog MA24106A UG Session Restore 17.Offset Table Menu Offset Table Off StatusUsing the Power Sensor Making Measurements Basic Power MeasurementConnecting the DUT Making MeasurementsCalibrating the Sensor Making Measurements Using the Power SensorZeroing the Sensor Applying a Calibration Factor CorrectionNumber Averages Needed for ±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm NoiseMeasurement Considerations Using the Power Sensor Error StatesTime Varying Signals Message Description ResolutionUsing the Power Sensor Measurement Considerations High Crest Factor Signals peak to average ratioMultitone Signals Noise and AveragingSettling Time Using the Power Sensor Uncertainty of a Measurement Uncertainty of a MeasurementUncertainty Example Uncertainty of a Measurement Using the Power SensorUncertainty Adjusted Specification Noise Calculations at 12 dBm 16 mWCommand Description Remote Operation Commands SummaryRemote Operation Remote Operation Command Details Remote Operation Command DetailsRemote Operation Remote Operation Command Details FREQ?HAT Sensor Operational Tests Required EquipmentVswr Pretest Test ProcedureVswr Pretest Sensor Operational Tests Calibration Factor Test Sensor Operational Tests Calibration Factor TestCalibration Factor Test Sensor Operational Tests MA24106ALinearity Test Sensor Operational Tests Linearity TestLinearity Test Sensor Operational Tests ApplySensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Appendix a Connector Care and Handling Beware of destructive pin depth mating connectorsCleaning connectors Avoid over torquing connectorsAvoid applying excessive power Connector Care and HandlingAppendix B Sample Visual Basic Code Using the Demo ApplicationDemo Application Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine For Anritsu Power Meter 1.0 Only Appendix C Serial Port CompatibilityMethod 1-Download Updated Software Method 3-Remapping a Serial PortSerial Port Compatibility Figure C-2.System PropertiesMethod 3-Remapping a Serial Port Figure C-3.Device ManagerFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Appendix D Upgrading the Firmware Upgrading the FirmwareFigure D-3.Firmware Upgrade Dialog Upgrading the FirmwareFigure D-5.Open File Dialog Figure D-7 Symbols to L SymbolsIndex To S Index-2 MA24106A UGTo Z USB