Remote Operation

Remote Operation Command Details

FREQ?

Description: Gets the current cal factor frequency value from the sensor.

Syntax: FREQ? + LF

Return Value: Current cal factor frequency in GHz

FREQ

Description: Sets the current cal factor frequency value for the sensor.

Syntax: FREQ fghz + LF

Return Value: OK or ERR

Remarks: “fghz” is the cal factor frequency value in GHz. “fghz” must be between 0.05 GHz to 6 GHz. Available resolution is 3 digits after the decimal point.

ZERO

Description: Zeros the power sensor. Syntax: ZERO + LF

Return Value: OK if successful and ERR if zeroing failed.

Remarks: In case of zero failure, the STATUS? command can be used to retrieve more detail about the error. See the STATUS? command for more detail. It takes 19 seconds to zero a sensor, please wait at least this long to get a response from the sensor.

TMP?

Description: Gets the current temperature reading from the sensor.

Syntax: TMP? + LF

Return Value: Current temperature in degrees C.

STATUS?

Description: Get error status byte from the sensor.

Syntax: STATUS? + LF

Return Value: Error status byte

Remarks: Error status byte information:

Status.b0 -> ZERO_TEMP_ERROR (Temperature changed more than allowable limit after zeroing sensor)

Status.b1 -> Not Used

Status.b2 -> ADC_CH2_OR (Temperature over range)

Status.b3 -> ADC_CH3_OR (Detector A over ranged)

Status.b4 -> ZERO_ERROR_DET_A

Status.b5 -> ZERO_ERROR_DET_B

Status.b6 -> TEMP_ERROR (Temperature beyond operating range)

Status.b7 -> Not Used

MA24106A UG

4-3

Image 39
Contents True-RMS, 50 MHz to 6 GHz MA24106A USB Power Sensor User GuideUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentInstallation Driver InstallationHardware and Software Requirements Anritsu Power Meter Installation Driver InstallationLicense Agreement Installation Driver InstallationSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Graphical User Interface Using the Power SensorButtons Using the Power Sensor ButtonsData Entry Fields Using the Power Sensor Display Window Display WindowStatus Bar Tools Menu Menu Bar Using the Power SensorMenu Bar File MenuDataLogging Menu Data Logging Menu10. Data Logging Power GraphTime, X-axis 30 minutes 14.Offset Entry Screen Offset Table16. Open Dialog MA24106A UG 17.Offset Table Menu Offset Table Off Status Session RestoreMaking Measurements Basic Power MeasurementConnecting the DUT Using the Power Sensor Making MeasurementsApplying a Calibration Factor Correction Making Measurements Using the Power SensorZeroing the Sensor Calibrating the Sensor±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm Noise Number Averages Needed forMessage Description Resolution Error StatesTime Varying Signals Measurement Considerations Using the Power SensorNoise and Averaging High Crest Factor Signals peak to average ratioMultitone Signals Using the Power Sensor Measurement ConsiderationsSettling Time Uncertainty of a Measurement Using the Power Sensor Uncertainty of a MeasurementNoise Calculations at 12 dBm 16 mW Uncertainty of a Measurement Using the Power SensorUncertainty Adjusted Specification Uncertainty ExampleRemote Operation Commands Summary Command DescriptionRemote Operation Remote Operation Command Details Remote Operation Command DetailsFREQ? Remote Operation Remote Operation Command DetailsHAT Required Equipment Sensor Operational TestsTest Procedure Vswr PretestVswr Pretest Sensor Operational Tests Sensor Operational Tests Calibration Factor Test Calibration Factor TestMA24106A Calibration Factor Test Sensor Operational TestsSensor Operational Tests Linearity Test Linearity TestApply Linearity Test Sensor Operational TestsSensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Beware of destructive pin depth mating connectors Appendix a Connector Care and HandlingConnector Care and Handling Avoid over torquing connectorsAvoid applying excessive power Cleaning connectorsUsing the Demo Application Appendix B Sample Visual Basic CodeDemo Application Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine Appendix C Serial Port Compatibility For Anritsu Power Meter 1.0 OnlyMethod 3-Remapping a Serial Port Method 1-Download Updated SoftwareFigure C-2.System Properties Serial Port CompatibilityFigure C-3.Device Manager Method 3-Remapping a Serial PortFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Upgrading the Firmware Appendix D Upgrading the FirmwareUpgrading the Firmware Figure D-3.Firmware Upgrade DialogFigure D-5.Open File Dialog Figure D-7 Symbols Symbols to LIndex Index-2 MA24106A UG To SUSB To Z