Sensor Operational Tests

Calibration Factor Test

5-4 Calibration Factor Test

In this test the calibration factors of the MA24106A are compared against another sensor (referred to in this procedure as the “reference sensor”) with known calibration factor uncertainties. This reference sensor should be calibrated by a reputable standards laboratory which has low published measurement uncertainty. To perform the comparison, both sensors are used to measure the output power of a synthesizer with a high quality attenuator, such as the 41KC-6, on the output. The attenuator improves the source match of the synthesizer by lowering the mismatch ripples, thereby lowering the uncertainty in the comparison.

Test Procedure

1.Set up the equipment as follows (refer to Figure 5-1 for an illustration):

a.Connect the reference power sensor to the reference power meter using the appropriate cables.

b.Connect the MA24106A USB cable between the personal computer with the Power Meter application installed and the MA24106A power sensor under test.

c.Launch the Power Meter application.

d.Turn the power on to all of the instruments and allow them to warm up for the amount of time specified in their respective manuals.

e.Reset or Preset all of the instruments.

f.Configure the reference meter and sensor to measure a CW signal.

g.Perform a sensor Zero and a 1 mW reference calibration on the reference sensor and meter per the manufacturer’s instructions.

h.Perform a low level Zero of the MA24106A as follows:

With the MA24106A disconnected from the synthesizer, click the Zero Sensor button on the Power Meter application and wait for the Zeroing Sensor message to close.

i.Connect the attenuator to the output of the synthesizer with the appropriate adapter to the output of the attenuator.

j.Set the synthesizer to +6 dBm and 50 MHz.

Figure 5-1.Cal Factor Test Set Up

MA24106A UG

5-3

Image 43
Contents True-RMS, 50 MHz to 6 GHz MA24106A USB Power Sensor User GuideUpdates Anritsu Company Software License Agreement Page Table of Contents Table of Contents General Information Preparation for Storage/Shipment Preparation for Storage/ShipmentDriver Installation InstallationHardware and Software Requirements Anritsu Power Meter Installation Driver InstallationLicense Agreement Installation Driver InstallationSoftware will then install to the selected location Found New Hardware Wizard Select the hardware type Computer, and then click Next Click Have Disk..., and then click Next 14.Found New Hardware Wizard Click Finish to close the wizard Page Graphical User Interface Using the Power SensorButtons Buttons Using the Power SensorData Entry Fields Display Window Using the Power Sensor Display WindowStatus Bar Tools Menu Menu Bar Using the Power SensorMenu Bar File MenuDataLogging Menu Data Logging Menu10. Data Logging Power GraphTime, X-axis 30 minutes 14.Offset Entry Screen Offset Table16. Open Dialog MA24106A UG 17.Offset Table Menu Offset Table Off Status Session RestoreMaking Measurements Basic Power MeasurementConnecting the DUT Using the Power Sensor Making MeasurementsApplying a Calibration Factor Correction Making Measurements Using the Power SensorZeroing the Sensor Calibrating the Sensor±0.20 dB ±0.15 dB ±0.10 dB ±0.05 dB ±0.01 dB DBm Noise Number Averages Needed forMessage Description Resolution Error StatesTime Varying Signals Measurement Considerations Using the Power SensorNoise and Averaging High Crest Factor Signals peak to average ratioMultitone Signals Using the Power Sensor Measurement ConsiderationsSettling Time Uncertainty of a Measurement Using the Power Sensor Uncertainty of a MeasurementNoise Calculations at 12 dBm 16 mW Uncertainty of a Measurement Using the Power SensorUncertainty Adjusted Specification Uncertainty ExampleCommand Description Remote Operation Commands SummaryRemote Operation Remote Operation Command Details Remote Operation Command DetailsFREQ? Remote Operation Remote Operation Command DetailsHAT Required Equipment Sensor Operational TestsVswr Pretest Test ProcedureVswr Pretest Sensor Operational Tests Sensor Operational Tests Calibration Factor Test Calibration Factor TestMA24106A Calibration Factor Test Sensor Operational TestsSensor Operational Tests Linearity Test Linearity TestApply Linearity Test Sensor Operational TestsSensor Operational Tests Linearity Test +20 +15 +10 Adjust per Max Min Delta E11 E12 Measurement Results 2 GHz Measurement Results 6 GHz Beware of destructive pin depth mating connectors Appendix a Connector Care and HandlingConnector Care and Handling Avoid over torquing connectorsAvoid applying excessive power Cleaning connectorsAppendix B Sample Visual Basic Code Using the Demo ApplicationDemo Application Using the Demo Application Sample Visual Basic Code Sample Visual Basic Code Using the Demo Application Delay routine Appendix C Serial Port Compatibility For Anritsu Power Meter 1.0 OnlyMethod 3-Remapping a Serial Port Method 1-Download Updated SoftwareFigure C-2.System Properties Serial Port CompatibilityFigure C-3.Device Manager Method 3-Remapping a Serial PortFigure C-4.Advanced Settings for COM Port Close all windows that you have opened up to this point Upgrading the Firmware Appendix D Upgrading the FirmwareUpgrading the Firmware Figure D-3.Firmware Upgrade DialogFigure D-5.Open File Dialog Figure D-7 Symbols to L SymbolsIndex Index-2 MA24106A UG To SUSB To Z