Agilent Technologies 8722ET, ES, 8720ET, 8719ET Meet Electrical and Environmental Requirements

Page 14

Installing Your Analyzer

STEP 3. Meet Electrical and Environmental Requirements

5.Ensure there are at least six inches of clearance between the sides and back of either the stand-alone analyzer or the system cabinet.

CAUTION The environmental temperature must be 4 °C less than the maximum operating temperature of the analyzer for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is >800 watts, then you must provide forced convection.

6.Set up a static-safe workstation. Electrostatic discharge (ESD) can damage or destroy electronic components.

static-control table mat and earth ground wire: part number 9300-0797

wrist-strap cord: part number 9300-0980

wrist-strap: part number 9300-1367

heel-straps: part number 9300-1308

• floor mat: not available through Agilent Technologies

1-8

Chapter 1

Image 14
Contents Installation and Quick Start Guide Part Number Print Date FebruaryPage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Familiarize Yourself with the Analyzer Front and Rear Panels Analyzer Front PanelAnalyzer Rear Panel Test SequenceMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test Service Menu Tests To Run the Operators CheckTo Test the Transmission Mode TransmissnTo Test the Reflection Mode Meas Refl REV S22 B/RDescription Eeprom Backup Disk ProcedureBack Up the Eeprom Disk EquipmentBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Measure the device Measurement ProcedureMake a measurement calibration Output measurement resultsLearning to Make Transmission Measurements Device Connections for a Transmission MeasurementPerform a measurement calibration Connections for a Thru Calibration StandardExample Measurement of Insertion Loss Press Return Define DISK-SAVE Measuring Other Transmission Characteristics Example Measurement of 3 dB BandwidthMeasuring Out-of-Band Rejection Example Measurement of Out-of-Band RejectionMeasuring Passband Flatness or Ripple Example Measurement of Passband Flatness or Ripple= 1 + ρ / 1 − ρ Learning to Make Reflection Measurements= Γ Connections for Reflection Measurements Make a measurement calibration Measuring Return Loss 10 Example Measurement of Return Loss11 Example Measurement Title Measuring Other Reflection Characteristics Measuring Reflection CoefficientMeasuring Standing Wave Ratio SWR 13 Example Standing-Wave-Ratio Measurement TraceMeasuring S11 and S22 or Reflection in a Polar Format Measuring S11 and S22 or Reflection in a Smith Chart Format 15 Example Impedance Measurement Trace If the analyzer display does not light If You Encounter a ProblemPower-Up Problems No RF Output Data Entry ProblemsIf there is no RF signal at the front-panel port If You Encounter a Problem Agilent Technologies Sales and Service Offices European Field OperationsIndex Index
Related manuals
Manual 63 pages 28.88 Kb