Agilent Technologies 8719ET, ES, 8720ET, 8722ET quick start Connections for Reflection Measurements

Page 47

Quick Start: Learning How to Make Measurements

Learning to Make Reflection Measurements

Step 1. Choose measurement parameters with your test device connected

1.Press the Preset key to return the analyzer to a known state.

2.Connect your test device as shown in Figure 2-8.If using a load, make sure it has the correct characteristic impedance.

CAUTION Damage may result to the device under test if it is sensitive to the analyzer's default output power level. To avoid damaging a sensitive DUT, be sure to set the analyzer's output power to an appropriate level before connecting the DUT to the analyzer.

Figure 2-8 Connections for Reflection Measurements

3.Choose the following measurement parameters:

Meas Refl: FWD S11 (A/R) or REFLECTION

Center10.24 G/n

Span1G/n

Scale Ref AUTO SCALE

Avg AVERAGING FACTOR 32 x1 . Toggle AVERAGING on OFF to ON.

4.Look at the device response to determine if these are the measurement parameters that you want. For example, if the trace is noisy, you may want to increase the input power, reduce the IF bandwidth, or add averaging. To better see an area of interest, change the test frequencies.

2-14

Chapter 2

Image 47
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTransmissn To Test the Transmission ModeMeas Refl REV S22 B/R To Test the Reflection ModeEquipment Eeprom Backup Disk ProcedureBack Up the Eeprom Disk DescriptionBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Output measurement results Measurement ProcedureMake a measurement calibration Measure the deviceDevice Connections for a Transmission Measurement Learning to Make Transmission MeasurementsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsExample Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or Ripple= 1 + ρ / 1 − ρ Learning to Make Reflection Measurements= Γ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring S11 and S22 or Reflection in a Smith Chart Format 15 Example Impedance Measurement Trace If the analyzer display does not light If You Encounter a ProblemPower-Up Problems No RF Output Data Entry ProblemsIf there is no RF signal at the front-panel port If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index