Agilent Technologies 8720ET, ES, 8722ET, 8719ET quick start Data Entry Problems, No RF Output

Page 57

Quick Start: Learning How to Make Measurements

If You Encounter a Problem

If the display lights, but the ventilation fan does not start:

Check that the fan is not obstructed. To check the fan, follow these steps:

1.Switch the LINE power to the off position.

2.Check that the fan blades are not jammed.

Contact the nearest Agilent Technologies office for service, if necessary. A list of Agilent Technologies sales and service offices is provided in Table 2-1 on page 2-26.

Data Entry Problems

If the data entry controls (keypad, knob, arrow keys) do not respond:

Check that the ENTRY OFF function is not enabled.

The ENTRY OFF function is enabled after you press the Entry Off key. To return to normal entry mode, press any function key that has a numeric parameter associated with it, for example, Start.

Check that none of the keys are stuck.

Check that the selected function key accepts data.

For example, Scale Ref accepts data, but System does not.

Check that the analyzer's "R" GPIB STATUS light is not illuminated.

If the analyzer's "R" GPIB STATUS light is illuminated, a test sequence may be running, or a connected computer controller may be sending commands or instructions

to, or receiving data from, the analyzer. Press Local if you want to return to LOCAL control.

If the parameter you are trying to enter is not accepted by the analyzer:

Ensure that you are not attempting to set the parameter greater than or less than its limit. Refer to the User's Guide for parameter limits.

No RF Output

If there is no RF signal at the front-panel port:

Check that the signal at the test port is switched on. 1. Press Power and toggle SOURCE PWR to ON.

NOTE

On ES models, it is possible to set the source power to come from PORT 2

 

instead of PORT 1, so you must check the power at the correct port. With

 

factory preset, the power comes from PORT 1.

 

 

If you are applying external modulation (AM) to the analyzer, check the external modulating signal or external gate/trigger signals for problems.

2-24

Chapter 2

Image 57
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTransmissn To Test the Transmission ModeMeas Refl REV S22 B/R To Test the Reflection ModeBack Up the Eeprom Disk Eeprom Backup Disk ProcedureDescription EquipmentBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Make a measurement calibration Measurement ProcedureMeasure the device Output measurement resultsDevice Connections for a Transmission Measurement Learning to Make Transmission MeasurementsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsExample Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or RippleLearning to Make Reflection Measurements = Γ= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring S11 and S22 or Reflection in a Smith Chart Format 15 Example Impedance Measurement Trace If You Encounter a Problem Power-Up ProblemsIf the analyzer display does not light Data Entry Problems If there is no RF signal at the front-panel portNo RF Output If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index