Agilent Technologies 8719ET, ES, 8720ET, 8722ET quick start Configure the Analyzer

Page 15

Installing Your Analyzer

STEP 4. Configure the Analyzer

STEP 4. Configure the Analyzer

This step shows you how to set up your particular analyzer configuration.

standard configuration

Option 1D5 configuration − high stability frequency reference

printer or plotter configuration

rack-mount configuration

Chapter 1

1-9

Image 15
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTransmissn To Test the Transmission ModeMeas Refl REV S22 B/R To Test the Reflection ModeEquipment Eeprom Backup Disk ProcedureBack Up the Eeprom Disk DescriptionBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Output measurement results Measurement ProcedureMake a measurement calibration Measure the deviceDevice Connections for a Transmission Measurement Learning to Make Transmission MeasurementsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsExample Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or RippleLearning to Make Reflection Measurements = Γ= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring S11 and S22 or Reflection in a Smith Chart Format 15 Example Impedance Measurement Trace If You Encounter a Problem Power-Up ProblemsIf the analyzer display does not light Data Entry Problems If there is no RF signal at the front-panel portNo RF Output If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index