Main Menu (continued)
■CALIBRATION
The CALIBRATION function in this unit measures LCD panel characteristics from low to high brightness values and internal monitor processing handles CALIBRATION.
CALIBRATION does not rely on image quality settings since internal signals are used for a calibration. CALIBRATION in this unit is made at D65 color temperature and calibrations for other color temperatures are results calculated from this value.
■Equipment required for calibration
•Konica Minolta
•Konica Minolta
•
•Connect the
| (Straight) | ||||||
Pin number | Signal |
|
|
|
| Pin number | Signal |
|
|
|
|
|
|
|
|
1 | CD |
|
|
|
| 1 | N.C. |
|
|
|
|
|
|
|
|
2 | RXD |
|
|
|
| 2 | TXD |
|
|
|
| ||||
|
|
|
|
|
|
|
|
3 | TXD |
|
|
|
| 3 | RXD |
|
|
|
| ||||
|
|
|
|
|
|
|
|
4 | DTR |
|
|
|
| 4 | DSR |
|
|
|
| ||||
|
|
|
|
|
|
|
|
5 | GND |
|
|
|
| 5 | GND |
|
|
|
| ||||
|
|
|
|
|
|
|
|
6 | DSR |
|
|
|
| 6 | DTR |
|
|
|
| ||||
|
|
|
|
|
|
|
|
7 | RTS |
|
|
|
| 7 | CTS |
|
|
|
| ||||
|
|
|
|
|
|
|
|
8 | CTS |
|
|
|
| 8 | RTS |
|
|
|
| ||||
|
|
|
|
|
|
|
|
9 | GND |
|
|
|
| 9 | N.C. |
|
|
|
|
|
|
|
|
This unit
Konica Minolta
Konica Minolta
•Be sure to set the following calibration values on the
Calibration | W | R | G | B |
|
|
|
|
|
x | 0.3091 | 0.6759 | 0.2029 | 0.1495 |
|
|
|
|
|
y | 0.3429 | 0.3126 | 0.691 | 0.0587 |
|
|
|
|
|
Lv | 195.3 | 47.56 | 135.2 | 13.21 |
|
|
|
|
|
•Turn on this unit and perform adequate aging (about 1 hour) before starting calibration.
•Turn off the lights in the room and make sure that no external light can enter the standard measurement probe before starting the calibration. External light that enters the probe will prevent correct calibration of low brightness characteristics.
•LCD panel characteristics and instrument error in the display color analyzer may sometimes result in small differences in values after calibration.
In a fine tuning of the monitor, also set GAIN and BIAS for R, G and B in the COLOR TEMP. VAR mode.
•Do not apply the probe to a WFM/VECTOR display on the screen.
30