Printer settings

To print calibration data via the RS232 interface, press the # + menu keys to recall

the System Set-up Screen. Using the keys, select PRINT and press ent.

Select the parameter to be printed using the keys, (set-point temperature [SET], dry well temperature [BLK], reference temperature [REF], measure [RDG]). Use the numeric keys to set the period, in seconds, between each printout and press ent. Press ce to return to the Working Screen.

SYSTEM SETUP

LANGUAGE

TEMPERATURE UNIT

KEYBOARD

SERIAL PORT

PRINT

PRINT SETUP

Period: 0 sec

SET NO

BLK NO

REF NO

RDG NO.

Calibration

A PIN code protects the CALIBRATION function. Refer to 5 Calibration, page 47.

SYSTEM SETUP

CALIBRATION

ENTER CODE

ENTER CODE

20

K400 Issue No. 1

Image 30
Contents Druck DBC 150/650 Series General Electric Company. All rights reserved Maintenance SafetyPower isolation Technical AdviceSymbols Traceability and Calibration ReportSafety Instructions AbbreviationsTable of Contents CAL Electrical Input Specification DBC-TC models Standard Accessories DBC-TC series Optional Accessories DBC-TC seriesStandard Accessories DBC-TS series Optional Accessories DBC-TS seriesTS Temperature Source TC Temperature CalibratorIntroduction Automatic Calibration Parts IdentificationDBC-TS Installation Power SupplyTemperature Probes Important Note Use of screw connections for T/C wires Electrical Connections Connection of a Thermocouple sensorConnection of a 3-wire RTD Connection of a 4-wire RTDOperating procedure GeneralCalibration with external reference probe External probe connections OPERATI0N Numeric keys +/- keyKey Arrow keysMenu selection screen System Set-up ScreenWorking screen Help/Info screen Self-test screenFunction Modes Reading of the generated temperature Dual display generated temperature and an input test optionTemperature Reference Selection Internal reference probeExternal reference probe SEL RTD System Set-up Language setting Temperature unit settingDigital interface settings Key-padaudio signalPrinter settings CalibrationProtection code Testing Modes Direct mode 150.00 CStep mode Fixed step selection19.00 C Ramp mode 50.00 CPreset values mode Switch Test Measuring an InputThermocouple Testing 148.00 CSetting Cold Junction Compensation InternalCAL Mode DirectAbsolute SpanReading RTD Measurements Pt-100 Using the keys, select the required method and press ent Reading Current Measurements CAL Voltage Measurements high level CAL Volt Voltage Measurements low level Resistance Measurement DBC Series Serial Commands Protocol Serial CommunicationTS/ TC Read AaOM 1? Preset mode French German Italian Portuguese Spanish Write Par = floating point Calibration Important NoteCalibration of the standard Reference Probe Input REF RTD Calibration of the millivolt input ranges DBC-TC only Calibration of the Volt input range DBC-TC only Calibration of the mA input range DBC-TC onlyCalibration of the resistance Ohm input ranges DBC-TC only Fault Finding Maintenance Cleaning InstructionsTechnical Specification Standard SpecificationCommon specification Electrical Input Specifications DBC-TC models Input Range Resolution Accuracy Days Year